Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bone Device Common Name: Drill Guide DNP 3.3mm Labeling: DRILL GUIDE DNP 3.3 MM NON-STERILE LOT NB0311 QTY 1 DePuy P.O Box 988 Warsaw IN 46581

Recall number
Z-2212-2014
Recalling firm
Biomet, Inc.
Product
Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bone Device Common Name: Drill Guide DNP 3.3mm Labeling: DRILL GUIDE DNP 3.3 MM NON-STERILE LOT NB0311 QTY 1 DePuy P.O Box 988 Warsaw IN 46581
Status
Terminated
Initiated
2014-06-23
Posted
2014-08-13
Terminated
2015-07-14
Reason
The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck.
Root cause
Nonconforming Material/Component
Action
On June 23, 2014 Biomet issued recall notification to their consignees via letter. Notification included affected product issue and instructions for return were provided. Distributors who had affected hospitals in their regions were notified and provided with a copy of the notice that was sent to the hospital. Consignees are being contacted on an individual basis in an attempt to reconcile the product 100%. A response form was provided to all consignees in order for Biomet to track returns. Non-responders will be contacted on an individual basis.
Quantity
37
Distribution
The product was distributed domestically and to internationally. Domestically to CA, PA, IA, CO, NY, MD, NC,AL, AZ, IL, GA, VA. Netherlands, Argentina,
Lot, serial or model codes
Catalog Number: DNPDG, Lot Number: NB0311 and NB0611
510(k) numbers
["K041081","K052248"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register