FDA Medical Device Recalls (openFDA)

Spacelabs Healthcare, Incorporated: Spacelabs Medical Integrated ECG Cable are supplied as part of the start-up kits provided with some Spacelabs' monitors. The kit PNs are: 718-0017-10, 718-0017-11, 718-0017-12, and 718-0017-13. The cables are also offered as replacement parts under Spacelabs PNs: 700-0008-22, 700-0008-23, 700-0008-24, and 700-0008-25. The Spacelabs elance Vital Signs Monitor is a prescription device intended to be used by healthcare professionals in all areas of healthcare.

Recall number
Z-2211-2009
Recalling firm
Spacelabs Healthcare, Incorporated
Product
Spacelabs Medical Integrated ECG Cable are supplied as part of the start-up kits provided with some Spacelabs' monitors. The kit PNs are: 718-0017-10, 718-0017-11, 718-0017-12, and 718-0017-13. The cables are also offered as replacement parts under Spacelabs PNs: 700-0008-22, 700-0008-23, 700-0008-24, and 700-0008-25. The Spacelabs elance Vital Signs Monitor is a prescription device intended to be used by healthcare professionals in all areas of healthcare.
Status
Terminated
Initiated
2009-07-23
Posted
2009-09-11
Terminated
2011-07-21
Reason
The Spacelabs Medical Integrated ECG Cables, PN: 700-0008-22, 700-0008-23, 700-0008-24, and 700-0008-25 will not withstand the high voltage of a defibrillation shock.
Root cause
Process control
Action
Spacelabs Healthcare issued an "Urgent: Medical Device Recall" dated July 20, 2009. Consignees are being sent replacement ECG cables for each item ordered and are requested to replace the affected cables. Also, they are instructed to destroy and dispose of the recalled cables according to their own protocol or local regulations. For further information, contact Spacelabs Healthcare Technical Support at 1-800-522-7025, option 2 (US) or +86-512-87171888, extension 304 (outside US).
Quantity
1,395
Distribution
International Distribution -- Bangladesh, Turkey, Trinidad and Tobago, Romania, Guatemala, Algeria, Netherlands, Egypt, Chile, Venezuela, Spain, Colombia, India, the United Kingdom, Singapore, France and Ecuador.
Lot, serial or model codes
Part Numbers: 700-0008-22, 700-0008-23, 700-0008-24, and 700-0008-25; Date Codes: OSC 12/08, LHI 02/09, LHI 03/09, LHI 04/09, and LHI 5/09.   ALL OTHER DATE CODES FUNCTION NORMALLY AND ARE NOT SUBJECT TO THIS RECALL.
510(k) numbers
["K090556"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3010157426
Firm city
Issaquah
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register