FDA Medical Device Recalls (openFDA)
Maquet Inc.: Maquet Getinge Group Pediatric Arterial Cannula Maquet Cardiopulmonary AG Kehler Strabe 31 76437 Rastatt Germany Phone +49 7222 932-0 Fax: +49 7222 932-1888 info.cp@maquet.com www.maquet.com Used to introduce blood and sterile solution to the circulation system during extracorporeal circulation procedures
- Recall number
- Z-2210-2012
- Recalling firm
- Maquet Inc.
- Product
- Maquet Getinge Group Pediatric Arterial Cannula Maquet Cardiopulmonary AG Kehler Strabe 31 76437 Rastatt Germany Phone +49 7222 932-0 Fax: +49 7222 932-1888 info.cp@maquet.com www.maquet.com Used to introduce blood and sterile solution to the circulation system during extracorporeal circulation procedures
- Status
- Terminated
- Initiated
- 2012-05-29
- Posted
- 2012-08-15
- Terminated
- 2014-04-03
- Reason
- Certain pediatric arterial cannulas may have limited flexibility due to an increased length of the "hardened" cannula tip.
- Root cause
- Device Design
- Action
- Maquet issued Urgent Device 2, Recall letters/Urgent Product Recall Forms dated May 29, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their current stock and identify any affected product and remove from their inventory. Customers were instructed to contact the customer service department at 800-777-4222 (option #2, followed by option #2). Customers were also instructed to completed the enclsoed Recall Form to acknowledge receipt of the notification and to document the current status of the devices shipped to their facility. For questions regarding this recall call 973-709-7660.
- Quantity
- 20 devices US; 122 devices foreign
- Distribution
- Worldwide Distribution - USA including MI, MN and Internationally
- Lot, serial or model codes
- 510k023251 US Catalog Number 161412 Batch Number 92048923 Foreign Catalog number 161412 Batch number 92048923 Catalog number 161412 Batch number 92051578
- 510(k) numbers
- ["K023251"]
- PMA numbers
- not on file
- Product code
- NCP
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28