FDA Medical Device Recalls (openFDA)
Skeletal Kinetics, Llc: InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate bone void filler. Sterilized using irradiation. The single-use InjectionPlasty 1.0 Kit contains the necessary components for mixing and delivery of the bone void filler. The InjectionPlasty 1.0 sterile kit contains: Calcium Phosphate Powder, Dilute Sodium Silicate Liquid, a Mixing System, Targeting K-Wires and a Delivery Needle (Cannula). Model number 10801210. InjectionPlasty 1.0 is indicated to fill bony voids or gaps of the skeletal system
- Recall number
- Z-2209-2015
- Recalling firm
- Skeletal Kinetics, Llc
- Product
- InjectionPlasty 1.0 Bone Void Filler; an injectable and biocompatible calcium phosphate bone void filler. Sterilized using irradiation. The single-use InjectionPlasty 1.0 Kit contains the necessary components for mixing and delivery of the bone void filler. The InjectionPlasty 1.0 sterile kit contains: Calcium Phosphate Powder, Dilute Sodium Silicate Liquid, a Mixing System, Targeting K-Wires and a Delivery Needle (Cannula). Model number 10801210. InjectionPlasty 1.0 is indicated to fill bony voids or gaps of the skeletal system
- Status
- Terminated
- Initiated
- 2015-05-29
- Posted
- 2015-07-24
- Terminated
- 2015-07-24
- Reason
- Reports of venous uptake, DVT, pulmonary embolism in patients who were implanted with Injection Plasty Bone Void Filler.
- Root cause
- No Marketing Application
- Action
- Skeletal Kinetics sent letters to distributors by overnight mail with instructions to forward a Dear Doctor notification letter on May 14, 2015. Second letters dated May 29, 2015, advised that the firm is removing the product from the marker. Distributor letters asked that the Dear Doctor letter be forward within 24 hours of receipt of the letter. All Injection Plasty kits in their possession and those with customers should be retrieved and returned by Fed Ex. Dear Doctor letters advised that their distribution agent has been requested to return all unused kits. Questions may be directed to 408-366-5002 of duran@skeletalkinetics.com.
- Quantity
- 67 kits
- Distribution
- US Distribution including Puerto Rico.
- Lot, serial or model codes
- Model number: 10801210: Lot Numbers Expiration Dates 13022501 8/31/2015 13052305 11/30/2015 13120901 5/31/2016 14051901 11/30/2016 14100101 3/31/2017 15042701 10/13/2017
- 510(k) numbers
- ["K051123"]
- PMA numbers
- not on file
- Product code
- MQV
- Firm FEI
- 3003890476
- Firm city
- Cupertino
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28