Cincinnati Sub-Zero Products Inc: Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer.

Recall number
Z-2209-2013
Recalling firm
Cincinnati Sub-Zero Products Inc
Product
Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer.
Status
Terminated
Initiated
2013-09-03
Posted
2013-09-16
Terminated
2014-08-22
Reason
During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.
Root cause
Labeling Change Control
Action
Cincinnati Sub-Zero (CSZ) sent an Urgent Medical Device Field Action letter dated September 5, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact CSZ at 513-772-8810 or 1-800-989-7373 to request new versions of the manuals. At that time customers are instructed to provide the quantity and serial numbers of the affected devices that are at their facility. Upon receipt of the new versions of the manual, immediately collect and destroy/discard all the affectred devices Operation and Technical Manuals their facility is currentlyin possession of and replace the discarded manuals with the new versions of the manuals provided by CSZ. After the manuals have been discarded and replaced, customers were asked to complete and return the enclosed response form as soon as possible to acknowledge receipt of this notification and to inform CSZ that they have performed and completed the requested actions. Customers should return the form by fax to 513-772-9119 scan and email the information to B3manual@cszinc.com or mail to Cincinnati Sub-Zero, 12011 Mosteller Road, Cincinnati, Ohio 45241. Customers with questions were instructed to contact their Blanketrol III Model 233 field action coordinator at 1-800-989-7373 or 513-772-8810. For questions regarding this recall call 513-772-8810, ext 5295.
Quantity
115V - 396 units; 230V - 385 units
Distribution
Worldwide Distribution - USA including AK, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, , NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI & WV. Internationally to ARGENTINA, AUSTRALIA, AZERBAIJAN, BELGIUM, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, GERMANY, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, KOREA, KUWAIT, MEXICO, NEW ZEALAND, PERU, SAUDI ARABIA, SINGAPORE, SPAIN, TAHITI, THAILAND, THE NETHERLANDS, THE PHILIPPINES, UNITED ARAB EMERITES & UNITED KINGDOM.
Lot, serial or model codes
Part number: 86107/06107/6107 (115V), Serial #s: 112-3-02021, 112-3-02102 through 124-3-02783, 074-3-00388, 081-3-00489, 081-3-00493, 082-3-00511, 091-3-00935, 094-3-01261, 094-3-01265, 094-3-01272, 094-3-01304, 094-3-01308, 101-3-01384, 093-3-01045, 093-3-01046, 093-3-01069, 084-3-00775, 073-3-00315, 091-3-00918, 091-3-00919, 073-3-00279, 094-3-01262, 071-3-00183, 092-3-00983, 094-3-01255, 064-3-00114, 072-3-00219, 072-3-00222, 072-3-00230, 072-3-00258, 081-3-00499, 081-3-00500, 064-3-00116, 064-3-00117, 092-3-01021, 073-3-00305, 073-3-00314, 082-3-22572, 082-3-22573, 101-3-01380, 073-3-00317, 074-3-00341, 091-3-00932, 094-3-01310, 101-3-01374, 082-3-00559, 092-3-00983, 094-3-01255, 094-3-01305, 062-3-00029, 093-3-01048, 094-3-01240, 102-3-01504 & 094-3-01243.  Part number: 86207/6207 (230V) Serial #s: 112-3-01949, 114-3-02171 through 124-3-02764, 072-3-00247, 073-3-00266, 073-3-00282, 082-3-00564, 071-3-00132, 071-3-00152, 073-3-00309, 082-3-00533, 082-3-00534, 082-3-00565, 082-3-00568, 084-3-00768, 084-3-00770, 084-3-00777, 084-3-00778, 084-3-00779, 084-3-00780, 084-3-00781, 084-3-00787, 084-3-00788, 091-3-00836, 091-3-00877, 091-3-00878, 091-3-00902, 092-3-01007, 092-3-01008, 092-3-01011, 102-3-01423, 102-3-01426, 103-3-01592, 103-3-01593, 103-3-01594, 103-3-01595, 103-3-01601, 093-3-01072, 093-3-01073, 093-3-01076, 093-3-01078, 093-3-01084, 093-3-01085, 093-3-01086, 093-3-01087, 093-3-01088, 093-3-01104, 093-3-01105, 093-3-01106, 093-3-01107, 093-3-01108, 103-3-01573, 103-3-01574, 103-3-01575, 071-3-00178, 071-3-00180, 074-3-00393, 074-3-00394, 074-3-00395, 074-3-00396, 074-3-00404, 074-3-00405, 091-3-00904, 091-3-00905, 092-3-00939, 092-3-00948, 092-3-00949, 092-3-00951, 092-3-00962, 092-3-00963, 094-3-01178,094-3-01179, 094-3-01180, 094-3-01181, 094-3-01183, 084-3-00789, 084-3-00790, 084-3-00791, 084-3-00812, 084-3-00813, 084-3-00814, 084-3-00815 & 084-3-00816.
510(k) numbers
["K101589"]
PMA numbers
not on file
Product code
DWJ
Firm FEI
1516825
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register