Brainlab AG: VectorVision (VV) Sky Navigation Platform (19" Computer Rack); the 19" computer rack is a component of the VVsky Vario, BrainSUITE iMRI, BrainSUITE NET and BrainSUITE iCT systems; BrainLab AG, Kapellenstrasse 12, 85622 Feldkirchen, Germany Intended to be an intraoperative image guided localization system to enable minimally invasive surgery. Indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified to relative to a CT, CTA, X-ray, MR, MRA and ultrasound based model of the anatomy.

Recall number
Z-2209-2008
Recalling firm
Brainlab AG
Product
VectorVision (VV) Sky Navigation Platform (19" Computer Rack); the 19" computer rack is a component of the VVsky Vario, BrainSUITE iMRI, BrainSUITE NET and BrainSUITE iCT systems; BrainLab AG, Kapellenstrasse 12, 85622 Feldkirchen, Germany Intended to be an intraoperative image guided localization system to enable minimally invasive surgery. Indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified to relative to a CT, CTA, X-ray, MR, MRA and ultrasound based model of the anatomy.
Status
Terminated
Initiated
2007-11-12
Posted
2008-09-17
Terminated
2009-07-24
Reason
Diameter of cables used for installation are to small for the applied current. If an internal short circuit is produced the medical power supply will not shut down automatically and will continue to deliver current, which could result in overheating cables.
Root cause
Component design/selection
Action
BrainLAB sent Product Notification letters dated 11/12/07 to their VectorVision Sky end users, informing them of the possible overload of electrical cables in VVsky products with Data Billboards and/or microscope integrations and/or Device Connections Panels (Wall Panels). The users were requested to perform an inspection of the medical power supplies and cabling on a regular basis. If anything suspicious is noted, they were directed to shut down the entire system and contact BrainLAB immediately. BrainLAB contacted each account shortly after the sending of the letter to arrange for the installation of small fuses/circuit breakers according to the specific load limits per device, which will open the circuit in case of overload on faulty circuits, and to exchange the low voltage cables that directly connect the medical power supply from 1.5 mm2 to 2.5 mm2 thickness to withstand the full 20 amp current in case of a failure. Any questions were directed to BrainLAB's customer hotline at 1-800-597-5911.
Quantity
23 units
Distribution
Nationwide Distribution --- including states of Arizona, Connecticut t, Idaho, Massachusetts, Michigan, Minnesota, New York, Ohio, Pennsylvania, Texas, Virginia and Wisconsin.
Lot, serial or model codes
VVsky Art. No. 50945; all VVsky products with Data Billboards and/or microscope integrations and/or Device Connections Panels (Wall Panels)
510(k) numbers
["K023651"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3002619595
Firm city
Kirchheim B. Muenchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register