Beckman Coulter Inc.: Access Immunoassay Systems AFP QC, Part Number: 33219 The Access AFP QC is intended for monitoring system performance of immunoenzymatic procedures for the quantitative measurement of alpha-fetoprotein (AFP) using the Access Immunoassay Systems.

Recall number
Z-2208-2010
Recalling firm
Beckman Coulter Inc.
Product
Access Immunoassay Systems AFP QC, Part Number: 33219 The Access AFP QC is intended for monitoring system performance of immunoenzymatic procedures for the quantitative measurement of alpha-fetoprotein (AFP) using the Access Immunoassay Systems.
Status
Terminated
Initiated
2010-04-21
Posted
2010-08-11
Terminated
2012-05-07
Reason
The recall was initiated after Beckman Coulter confirmed an issue with the Access AFP QC kits (REF 33219) which contain three levels of control material identified as QC1, QC2, and QC3, Beckman Coulter has confirmed that vials of QC1 in the kit lots identified above contain microbial contamination. Customers may see that the results from a contaminated vial of QC1 are within the established refere
Root cause
Other
Action
A Product Corrective Action (PCA) letter was sent on the week of April 19, 2010 to all customers who purchased the Access Immunoassay Systems AFP QC material, lot numbers referenced above. The letter provided the customers with an explanation of the problem identified and instructed the customers to (1) Please discard all remaining inventory of the three Access AFP QC kit lots identified above, including all vials of QC1, QC2, and QC3. (2) Please review your ac and patient results obtained while using the affected control lots. Interpret Access AFP test results in light of the clinical presentation of the patient, including: symptoms, clinical history, data from other tests and other appropriate information. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. (4) Complete and return the enclosed response form within 10 days so we can be assured that you have received this important notification. For customers that need assistance or have any questions regarding this notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
Quantity
166
Distribution
Nationwide and Canada
Lot, serial or model codes
Lot Numbers: 911158, 911581, and 917631
510(k) numbers
["K981864"]
PMA numbers
not on file
Product code
JJX
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register