FDA Medical Device Recalls (openFDA)
Medtronic Neurosurgery: Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.
- Recall number
- Z-2207-2023
- Recalling firm
- Medtronic Neurosurgery
- Product
- Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.
- Status
- Open, Classified
- Initiated
- 2023-06-13
- Posted
- 2023-07-20
- Terminated
- not on file
- Reason
- Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection.
- Root cause
- Process change control
- Action
- On 6/13/23 recall notices were mailed to customers and distributors and they were asked to do the following: 1) Return affected product to the recalling firm. 2) Complete and return the Customer Confirmation Form to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Customers with questions can contact Customer Service at 800-874-5797 Option 1.
- Quantity
- 65029
- Distribution
- Worldwide - US Nationwide distribution including in the states of PR, TX, WA, HI, MI, GA, AL, IL, IN, IA, MO, KY, NY, CA, AZ, PA, UT, ME, MD, MN, FL, AK, NE, VA, OR, WI, NC, LA, SD, TN, MS, DE, NJ, ID, SC, OH, MA, DC, NM, WV, NH, CT, NV, OK, AR, KS, MT, RI, CO, ND, WY, VT and the countries of Barbados, Cayman Islands, Honduras, Belgium, Switzerland, Italy, United Kingdom, Canary Islands, Hungary, Poland, Germany, Spain, Czech Republic, Thailand, South Africa, Greece, Finland, Portugal, Colombia, Norway, Iraq, Jordan, Estonia, Austria, Slovakia, Libya, Qatar, Bosnia And Herzegovina, Algeria, Dominican Republic, Canada, Australia, New Zealand, Sweden, El Salvador, Ireland, United Arab Emirates, Saudi Arabia, Kazakhstan, Turkey, Mauritius, Ghana, Russian Federation, Lebanon, Romania, Iran, Islamic Republic Of, Egypt, Namibia, Senegal, Ecuador, Panama, Peru, Guyana, Mexico, Chile, Costa Rica, Viet Nam, Kuwait, Ukraine, Israel, Turkmenistan.
- Lot, serial or model codes
- REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046; 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033; 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302; 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040; 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016; 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020, 2109001, 2109020, 2109028, 2110014, 2110032, 2110033, 2111028, 2111035, 2111037, 2112001, 2112002, 2202038, 2203023, 2204001, 2204028, 2206013, 2206027, 2207017, 2207048, 2208009; 62105/January 31, 2025/00643169063860/2008005, 2008041, 2009029, 2010015, 2010020, 2012002, 2101016, 2102003, 2102013, 2102029, 2103006, 2103036, 2104044, 2105036, 2106035, 2108003, 2108015, 2109002, 2109003, 2110003, 2110015, 2110016, 2111022, 2111033, 2111047, 2201006, 2202014, 2202025, 2203015, 2203025, 2203031, 2204023, 2204036, 2204041, 2206014, 2206028, 2207018, 2207044; 62106/December 31, 2024/00643169063877/2008042, 2010021, 2101017, 2102023, 2103007, 2104045, 2105039, 2106007, 2108004, 2108021, 2109005, 2110017, 2110034, 2111029, 2202039, 2207019, 2207051; 62110/March 31, 2025/00643169063884/2006041, 2007022, 2008006, 2008043, 2010022, 2101009, 2101018, 2102026, 2102049, 2103020, 2104046, 2105001, 2105040, 2108005, 2108022, 2110004, 2110018, 2111036, 2201007, 2202015, 2202031, 2204017, 2204042, 2205031, 2205032, 2206007, 2207020, 2207049; 62111/May 31, 2025/00643169063891/2010023, 2012003, 2101019, 2102024, 2103008, 2105034, 2106009, 2108023, 2109006, 2110019, 2110035, 2111030, 2111034, 2202040, 2203024, 2204002, 2207021, 2207037, 2208027; 62158/April 30, 2025/00643169063907/2011019, 2103009, 2105035, 2110020, 2111031, 2202041, 2203033, 2204029, 2207006
- 510(k) numbers
- ["K161370"]
- PMA numbers
- not on file
- Product code
- GXQ
- Firm FEI
- 3015531529
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28