FDA Medical Device Recalls (openFDA)
Monarch Labs, LLC: Monarch Labs Medical Maggots with the accessory LeFlap dressing (previously known as Creature Comforts II), 250-500 per vial, Batch #: MM-090406 with CCII-0901. Maggot therapy is intended to debride serious non-healing (chronic) wounds, and the dressing intended to confine the maggots to the wound.
- Recall number
- Z-2207-2010
- Recalling firm
- Monarch Labs, LLC
- Product
- Monarch Labs Medical Maggots with the accessory LeFlap dressing (previously known as Creature Comforts II), 250-500 per vial, Batch #: MM-090406 with CCII-0901. Maggot therapy is intended to debride serious non-healing (chronic) wounds, and the dressing intended to confine the maggots to the wound.
- Status
- Terminated
- Initiated
- 2009-05-03
- Posted
- 2010-08-11
- Terminated
- 2012-12-13
- Reason
- Monarch Labs initiated the field notification regarding their Medical Maggots from Batch MM090406 with the accessory LeFlap dressing due a customer complaint from a physician of a patient developing cellulitis after using maggots from the affected batch.
- Root cause
- Other
- Action
- The customer recall communication was initiated on May 3, 2009 with Monarch Labs notifying all consignees with a via a faxed Recall Letter with fax-back attachment of the initial compliant they had received and the possible infection that may occur if instructions for use are not followed. All therapists, who used maggots from that batch were instructed to contact the firm immediately if they have experienced any adverse events, such as infection, cellulitis, or any other problem. If not, they were asked to simply return the fax-back attachment , noting "no problem" .
- Quantity
- 14 vials
- Distribution
- Nationwide Distribution -- CA, IA, KY, IL, TX, SC & UT.
- Lot, serial or model codes
- Batch #: MM-090406 with CCII-0901
- 510(k) numbers
- ["K072438"]
- PMA numbers
- not on file
- Product code
- NQK
- Firm FEI
- 3005735989
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28