FDA Medical Device Recalls (openFDA)
Biomet Microfixation, Inc.: BIOMET MICROFIXATION "1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, QTY 1, 1.5 X 4MM TITANIUM. REF 91-1504 1 COUNT and REF 95-1504 - 5 PACK. BIOMET MICROFIXATION 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA, WWW.BIOMETMICROFIXATION.COM. Intended to aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.
- Recall number
- Z-2206-2010
- Recalling firm
- Biomet Microfixation, Inc.
- Product
- BIOMET MICROFIXATION "1.5MM" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, QTY 1, 1.5 X 4MM TITANIUM. REF 91-1504 1 COUNT and REF 95-1504 - 5 PACK. BIOMET MICROFIXATION 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA, WWW.BIOMETMICROFIXATION.COM. Intended to aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.
- Status
- Terminated
- Initiated
- 2010-03-17
- Posted
- 2010-08-11
- Terminated
- 2010-08-11
- Reason
- Biomet Microfixation has received a complaint from a customer who reports finding a 3.5mm length screw in one 5 pack of 4mm length screws.
- Root cause
- Other
- Action
- Biomet Microfixation is recalling their 1.5x4mm HT X-drive Screws Lot Number 148530 after receiving a customer reported finding one 3.5MM screw in a package labeled as containing 4.0MM screws. Lot 148530 was packaged into one count packages 91-1504 and 5 count packages 95-1504. The 1 and 5 count packages were distributed between 12/10/2009 and 02/01/2010. Beginning on 03/17/2010 customers identified as receiving Lot 148530 HT X-drive Screws were notified of the recall by a combination of an Urgent Medical Device Recall Notice letter, e-mail, fax and telephone calls. Customers were instructed to inspect their inventory and to return any affected products. They were instructed to complete the attached Inventory Reconciliation sheet and to fax it to Biomet Microfixation. Customers were also instructed to contact any customers to whom they may have further distributed the product. Questions should be directed to Christy Cain at 1-800-874-7711 or 904-741-4400, ext. 468.
- Quantity
- 1247 screws
- Distribution
- Worldwide Distribution -- USA, including states of FL, GA, IN, LA, MD, MI, MO, NC, NY, OK, TX, and VA and countries of Philippines, Japan, Brazil, and Canada.
- Lot, serial or model codes
- Lot 148530
- 510(k) numbers
- ["K953385"]
- PMA numbers
- not on file
- Product code
- DZL
- Firm FEI
- 1032347
- Firm city
- Jacksonville
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as