Skip to the content

FDA Medical Device Recalls (openFDA)

Umano Medical, Inc.: AC Powered hospital adjustable bed.

Recall number
Z-2205-2020
Recalling firm
Umano Medical, Inc.
Product
AC Powered hospital adjustable bed.
Status
Terminated
Initiated
2020-05-04
Posted
not on file
Terminated
2020-12-30
Reason
Potential Hazard of Unsupervised Patient being out of Bed and/or Potential Patient Fall.
Root cause
Use error
Action
The firm sent a Medical Device Recall-User Manual letter dated May 4,2020 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to identify affected units. Units bearing the options numbers OL360011P or OL360017P and manufactured between January 26th, 2015 and January 24rd,2020 were packaged with a User Manual affected by this recall and subject to improvement with the attached addendum. A list of serial numbers traced as being in your facility is attached to this communication. The manufacturing date and serial number are located on the manufacturer label located at the head end of the device. Actions Needed: 1. Please carefully read the attached addendum and promptly disseminate its contents to all concerned personnel in your facility, including more importantly to all persons who are or may be called upon to use/operate the bed. 2. Locate the User Manual related to your FL36 bed and insert the addendum, for future reference, in the appropriate section of the instruction manual for the FL36 ook-snow, ook-snow MH or ook-cocoon beds. 3. If you have loaned or sold any of the beds listed in this letter, please, forward a copy of this notice to the new users and advise us of their new location. 4. If you have disposed of any of the affected beds and they are no longer in use, please, advise us of their obsolescence by providing us with their serial numbers. 5. Call at 1-844-409-4030 or at 1-418-247-3986, extension 4 to confirm that your facility received Umano Medical Inc.s addendum to the User Manual regarding the Zeroing Procedure and the Detection System applicable to the above-mentioned product(s) and that the Actions Needed have been taken and completed by your facility.
Quantity
3259 beds
Distribution
Domestic distribution: AR, CA, FL, GA, UT, MS, NY, PA International distribution: Australia, Brasil, Canada and United Kingdom.
Lot, serial or model codes
Model FL36 Numbers OL360011P or OL360017P
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNL
Firm FEI
3009591865
Firm city
L'Islet
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register