Skip to the content

FDA Medical Device Recalls (openFDA)

Agfa N.V.: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications

Recall number
Z-2202-2020
Recalling firm
Agfa N.V.
Product
AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications
Status
Terminated
Initiated
2020-04-06
Posted
not on file
Terminated
2023-04-03
Reason
Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Root cause
Software design
Action
Agfa issued Urgent Field Safety Notice via email on 4/21/20 to consignees, letter states reason for recall, health risk and action to take: Acknowledgment, via a feedback form or email, that the information was received and understood was requested from the consignees.
Quantity
13 units
Distribution
US Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.
Lot, serial or model codes
Serial Numbers: 0925, 0945, 0948, 0949, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, and 1404
510(k) numbers
["K180589"]
PMA numbers
not on file
Product code
JAA
Firm FEI
3001556265
Firm city
Mortsel
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register