Gambro Renal Products, Incorporated: Prismaflex M100. Product number 106697. Each set is packaged in a plastic pouch, 4 pouches per carton box

Recall number
Z-2202-2014
Recalling firm
Gambro Renal Products, Incorporated
Product
Prismaflex M100. Product number 106697. Each set is packaged in a plastic pouch, 4 pouches per carton box
Status
Terminated
Initiated
2014-07-18
Posted
2014-08-13
Terminated
2015-02-05
Reason
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root cause
Device Design
Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity
401,451
Distribution
Worldwide Distribution.
Lot, serial or model codes
Prismaflex M100. Product number 106697. Batch number 12A0401G to 14C2802G. Expiry 01/2014 to 03/2016
510(k) numbers
["K041005"]
PMA numbers
not on file
Product code
KDI
Firm FEI
1713683
Firm city
Lakewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register