Myoscience Inc: iovera 155 Smart Tip; Catalog numbers:STT0412-10 and STT0412-5 (10 and 5 count respectively) iovera Smart Tips are individually packaged in Tyvek PET pouches, which are then packaged into 5 or 10 pack boxes and shipper boxes. Product Usage: Cyrogenic Surgical Device The myoscience iovera device is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The iovera device is not indicated for treatment of central nervous system tissue

Recall number
Z-2201-2015
Recalling firm
Myoscience Inc
Product
iovera 155 Smart Tip; Catalog numbers:STT0412-10 and STT0412-5 (10 and 5 count respectively) iovera Smart Tips are individually packaged in Tyvek PET pouches, which are then packaged into 5 or 10 pack boxes and shipper boxes. Product Usage: Cyrogenic Surgical Device The myoscience iovera device is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The iovera device is not indicated for treatment of central nervous system tissue
Status
Terminated
Initiated
2015-07-01
Posted
2015-07-22
Terminated
2015-09-08
Reason
The expiration date on the outer box label and the pouch label for the iovera Smart Tip product is incorrect. Instead of indicating the correct expiration or 2015-12. the labels were incorrectly labeled 2016-05.
Root cause
Incorrect or no expiration date
Action
Myoscience sent an Urgent Medical Device Correction letters dated July 1, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are requested to quarantine any affected product at their site. Myoscience will make arrangements to either correct the incorrect expiation date label or remove the product and replace it with correctly labeled product immediately. Question or concerns may be directed to Customers Service at 510-933-1500.
Quantity
12 boxes - 3 boxes of 10 count and 9 boxes of 5 count.
Distribution
US Nationwide Distribution in the states of TX, NY, MD, CO and CA.
Lot, serial or model codes
Catalog numbers:STT0412-¿10 and STT0412-¿5;  Lot numbers : 1504021-¿01 and 1505008.
510(k) numbers
["K142866"]
PMA numbers
not on file
Product code
GXH
Firm FEI
3009131456
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register