Skip to the content

FDA Medical Device Recalls (openFDA)

Nico Corp.: 13.5 mm x 60 mm BrainPath Sheath The NICO BrainPath is a family of products that consists of multiple sized reusable and re-sterilizable obturators with coordinating single patient use, disposable sheaths (three lengths available for each: 50mm, 60mm, and 75mm). The obturator and sheath are assembled in the OR immediately prior to use. These two components are held together by an interference fit. The device when used as intended is designed to provide access to neurological tissues.

Recall number
Z-2201-2012
Recalling firm
Nico Corp.
Product
13.5 mm x 60 mm BrainPath Sheath The NICO BrainPath is a family of products that consists of multiple sized reusable and re-sterilizable obturators with coordinating single patient use, disposable sheaths (three lengths available for each: 50mm, 60mm, and 75mm). The obturator and sheath are assembled in the OR immediately prior to use. These two components are held together by an interference fit. The device when used as intended is designed to provide access to neurological tissues.
Status
Terminated
Initiated
2012-07-17
Posted
2012-08-13
Terminated
2012-11-07
Reason
BrainPath sheaths were mislabeled. Sheaths which measure 50mm were labeled as 60mm. The length callout on the label does not match the actual length of the sheath within the packaging resulting in an improper fit as further explained below. This mislabeling was discovered on 29-June-2012 during a clinical case.
Root cause
Labeling mix-ups
Action
NICO Neuro and Spine sent an Urgent Recall letter to all affected customers on July 18, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete the recall response form. Affected sheaths distributed within the US will be obtained and discarded by NICO. Affected sheaths shipped to Canada will be obtained and discarded by NICOs distributor, Galen Medical. For questions customers should call 888-632-7071.
Quantity
15 devices (3 boxes, 5 devices per box)
Distribution
Worldwide Distribution - USA including MI and Canada
Lot, serial or model codes
REF: NN-8012 - LOT: 80345-TD00214 (manufactured on 11-April-2012)
510(k) numbers
["K120691"]
PMA numbers
not on file
Product code
GZT
Firm FEI
3007456622
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register