FDA Medical Device Recalls (openFDA)
Philips Ultrasound Inc: Philips EPIQ and Affiniti Ultrasound Systems with software version 4.0: Models EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIZ CVx, Affiniti 30, Affiniti 50, and Affiniti 70. Product Usage: The intended use of the EPIQ, EPIQ 5, EPIQ 7, Affiniti 30, Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Or Jan (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal. When integrated with Philips EchoNavigator, the systems can assist the interventionist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.
- Recall number
- Z-2199-2019
- Recalling firm
- Philips Ultrasound Inc
- Product
- Philips EPIQ and Affiniti Ultrasound Systems with software version 4.0: Models EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIZ CVx, Affiniti 30, Affiniti 50, and Affiniti 70. Product Usage: The intended use of the EPIQ, EPIQ 5, EPIQ 7, Affiniti 30, Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Or Jan (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal. When integrated with Philips EchoNavigator, the systems can assist the interventionist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.
- Status
- Terminated
- Initiated
- 2019-04-29
- Posted
- 2019-08-06
- Terminated
- 2022-08-04
- Reason
- There is a potential that the image of one patient could get unexpectedly moved into another patient's folder when the user is using the Edit functionality on the Patient Data Entry (PDE) screen of the device.
- Root cause
- Software design
- Action
- The recalling firm issued letters dated 4/26/2019 on 4/29/2019 via certified mail with delivery confirmation. The letter identified the affected product, problem and actions to be taken.
- Quantity
- 3,146 devices
- Distribution
- Worldwide Distribution - US Nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Latvia, Luxembourg, Malaysia, Mayotte, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Norway, Oman, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United Kingdom Virgin Islands, Uruguay, and Viet Nam.
- Lot, serial or model codes
- All models containing software version 4.0.
- 510(k) numbers
- ["K182857"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 3019216
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28