Skip to the content

FDA Medical Device Recalls (openFDA)

Capnia Inc: Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Recall number
Z-2198-2025
Recalling firm
Capnia Inc
Product
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Status
Open, Classified
Initiated
2025-04-21
Posted
2025-07-29
Terminated
not on file
Reason
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Root cause
Device Design
Action
Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986
Quantity
2160
Distribution
US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
Lot, serial or model codes
Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
510(k) numbers
["K151107"]
PMA numbers
not on file
Product code
CCJ
Firm FEI
3008614998
Firm city
Foster City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register