Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.

Recall number
Z-2198-2016
Recalling firm
Synthes (USA) Products LLC
Product
Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
Status
Terminated
Initiated
2016-05-31
Posted
2016-07-12
Terminated
2017-02-02
Reason
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
Root cause
Labeling False and Misleading
Action
DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
Quantity
36 units
Distribution
US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
Lot, serial or model codes
Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWE
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register