FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
- Recall number
- Z-2198-2016
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.
- Status
- Terminated
- Initiated
- 2016-05-31
- Posted
- 2016-07-12
- Terminated
- 2017-02-02
- Reason
- DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Root cause
- Labeling False and Misleading
- Action
- DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
- Quantity
- 36 units
- Distribution
- US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
- Lot, serial or model codes
- Part # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWE
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as