FDA Medical Device Recalls (openFDA)
Assuramed: ProTime Microcoagulation System; PROTIME CUVETTE AND PROTIME3 CUVETTE; Store the foil-pouched cuvettes refrigerated 2-8C (36 - 36F); MATERIALS PROVIDED; ProTime cuvettes with Tenderlett Plus; or; Protime3 cuvettes with Tenderlett Plus LV; Product Instructions; Antiseptic, gauze (with self-test package only); International Technidyne Corporation; 8 Olsen Avenue; Edison, NJ 08820 USA; a subsidiary of Thoratec Corporation The ProTime Microcoagulation System consists of a portable, battery operated instrument and disposable cuvette for quantitative determination of prothrombin time (PT) from fingerstick whole blood or anticoagulant-free venous whole blood. There are two different types of cuvettes available for use with the ProTime Microcoagulation System: a ProTime cuvette and a ProTime3 cuvette. The ProTime cuvette is black color coded and utilizes three channels for the PT assays and two channels for integral controls. It requires approximately 65 ul of blood (approximately 3 drops). The ProTime3 cuvette is blue color coded and utilizes one channel for PT assay and two channels for integral controls. The ProTime3 cuvette requires approximately 27 ul of blood (approximately 1 large drop). The ProTime Microcoagulation System is intended for professional use in the management of patients treated with oral anticoagulants or for patient self-testing.
- Recall number
- Z-2198-2013
- Recalling firm
- Assuramed
- Product
- ProTime Microcoagulation System; PROTIME CUVETTE AND PROTIME3 CUVETTE; Store the foil-pouched cuvettes refrigerated 2-8C (36 - 36F); MATERIALS PROVIDED; ProTime cuvettes with Tenderlett Plus; or; Protime3 cuvettes with Tenderlett Plus LV; Product Instructions; Antiseptic, gauze (with self-test package only); International Technidyne Corporation; 8 Olsen Avenue; Edison, NJ 08820 USA; a subsidiary of Thoratec Corporation The ProTime Microcoagulation System consists of a portable, battery operated instrument and disposable cuvette for quantitative determination of prothrombin time (PT) from fingerstick whole blood or anticoagulant-free venous whole blood. There are two different types of cuvettes available for use with the ProTime Microcoagulation System: a ProTime cuvette and a ProTime3 cuvette. The ProTime cuvette is black color coded and utilizes three channels for the PT assays and two channels for integral controls. It requires approximately 65 ul of blood (approximately 3 drops). The ProTime3 cuvette is blue color coded and utilizes one channel for PT assay and two channels for integral controls. The ProTime3 cuvette requires approximately 27 ul of blood (approximately 1 large drop). The ProTime Microcoagulation System is intended for professional use in the management of patients treated with oral anticoagulants or for patient self-testing.
- Status
- Terminated
- Initiated
- 2013-07-16
- Posted
- 2013-09-13
- Terminated
- 2017-12-27
- Reason
- The products may have been stored at temperatures outside their required storage conditions.
- Root cause
- Storage
- Action
- Independence Medical (im) sent an Urgent Product Recall letter dated July 16, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to identify and immediately cease the distribution of the recalled products, quarantine the recalled products, and contact the recalling firm at 800-860-8027 for instructions on how to return the recalled products. Consignees who further distributed the recalled products were requested to notify their customers of the recall. Consignees were also instructed to complete the Product Recall Acknowledgement Form and return the document by email to recall@indemed.com indicating receipt of the notification.Customers with questions were instructed to contact the product manufacturer for ProTime Disposable Cuvettes at 800-579-2255, DiaScreen Liquid Urine Controls at 800-818-8877, and Hemoccult II SENSA elite Dispensapak at 800-854-3633. For questions regarding this recall call 800-860-8027.
- Quantity
- 1) 72 units; 2) 56 units
- Distribution
- Nationwide Distribution including AR, AZ, CA, CO, FL, IL, KY, LA, MI, NJ, NY, OH,OK, OR, SC, TX, UT, and WA.
- Lot, serial or model codes
- *** 1) Product Number: PRO3-25; Lots: All Lots; Independence Medical Catalog Number: 55L2101002; *** 2) Product Number: PRO5-25; Lots: All Lots; Independence Medical Catalog Number: ICPRO525
- 510(k) numbers
- ["K010599"]
- PMA numbers
- not on file
- Product code
- GJS
- Firm FEI
- 3009929723
- Firm city
- Twinsburg
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28