Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer

Recall number
Z-2197-2020
Recalling firm
Olympus Corporation of the Americas
Product
OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer
Status
Terminated
Initiated
2020-04-23
Posted
not on file
Terminated
2024-10-09
Reason
A component of the Olympus ShockPulse Transducer was manufactured with non-conforming material that could allow water ingress during autoclaving sterilization.
Root cause
Component design/selection
Action
Olympus sent a Medical Device Field Removal Action letter, dated April 23, 2020, to all ShockPulse Transducer (Model SPL-T) customers with the affected serial numbers. The letter states that the transducer was manufactured with non-conforming material and could become warped during sterilization. They explained that the serial number of the transducers is found etched on the back of the device. The firm asked that the customers determine whether they have any of the affected product in stock. If customers have any of the affected product , they were instructed to quarantine the product. and contact the Olympus customer service representative at 1-888-524-7266 option 1, in order to receive a Return Material Authorization for any affected product then, get issued a credit or replacement. Customers were also instructed to access the Olympus recall portal (https://olympusamerica.com/recall) to indicate that they received the notification. If they distributed the product further, the firm asked them to notify their customers to inform them of the field action.
Quantity
18
Distribution
Distribution US nationwide to states of: Texas, Ohio, New Mexico, California, Oklahoma, Michigan, Connecticut, North Carolina, Nebraska, and Pennsylvania; and Canada.
Lot, serial or model codes
Serial numbers: 1910PW30607705, 1910PW30607712, 1910PW30607601, 1910PW30607708, 1910PW30607709, 1910PW30607701, 1910PW30607814, 1910PW30607906, 1910PW30607603, 1910PW30607706, 1910PW30607802, 1910PW30607808, 1910PW30607809, 1910PW30607804, 1910PW30607713, 1910PW30607803  Lot Numbers: CG2003, CG2010, CG2011, CG2026, CG2027, CG2033, CG2035, CG2000, CG2025, CG2001, CG2019, CG2022, CG2029, CG2021, CG2008, CG2036, CG2037, CG2006, CG2004, CG2007, CG2012, CG2013, CG2015
510(k) numbers
["K142428"]
PMA numbers
not on file
Product code
FEO
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register