Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650

Recall number
Z-2195-2010
Recalling firm
Beckman Coulter Inc.
Product
IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
Status
Terminated
Initiated
2010-05-21
Posted
2010-08-10
Terminated
2012-05-22
Reason
Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have been reported to cause high shifts in recovery of control or patient samples. Quality Control failures may result after replacing BUF1 with a different lot.
Root cause
Nonconforming Material/Component
Action
Beckman Coulter will send a Product Corrective Action (PCA) letter on the week of April 19, 2010 to the affected customers. The letter provided the customers with an explanation of the problem and a work around.
Quantity
1309
Distribution
Worldwide: U.S., Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Iceland, Ireland, Italy, Japan, Liechtenstein, Macao, Malaysia, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey and United Kingdom,
Lot, serial or model codes
Lot#s: T908069, T908104, T909110, T001036 ( This lot was not shipped to customers.)
510(k) numbers
["K962294"]
PMA numbers
not on file
Product code
JQX
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register