Philips Medical Systems (Cleveland) Inc: Vereos PET/CT Model # 882446, computed tomography x-ray system

Recall number
Z-2194-2019
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Vereos PET/CT Model # 882446, computed tomography x-ray system
Status
Terminated
Initiated
2019-06-24
Posted
2019-08-06
Terminated
2022-10-06
Reason
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Root cause
Other
Action
Philips Healthcare issued a Field Safety Notification (FSN)72800720/88200522 1. Identify affected product 2. There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images. A. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs when Start Final Recon is selected prior to display of the ECG wave at the top of the acquisition window. B. When performing a Step and Shoot acquisition with phase tolerance selected, images may be reconstructed identically (a single phase), but are labeled as different phases. This occurs if ECG leads become disconnected mid-acquisition or when the acquisition is halted prematurely due to an application crash. C. By system design, in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase 4. Determine whether you have a potentially affected system, then determine your AFFECTED PRODUCTS software version. To identify the software version of your product: Click the Help button, Select About and the software version is then displayed, Software versions 4.x, 3.x, and 2.x are potentially affected. 5. Take the Action listed in the field safety notice. Please retain a copy with the equipment Instruction for use. If you need any further information or support concerning this issue, please contact your local Philips representative: North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377 and follow the prompts).
Quantity
26
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, and Yemen
Lot, serial or model codes
System Serial Number 900004 through 900076
510(k) numbers
["K123599"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register