FDA Medical Device Recalls (openFDA)
Carefusion 2200 Inc: *** 1) Cat. 4341B; Qty. 10; Thoracentesis Tray; With Catheter; Sterile; CareFusion; 1500 Waukegan Road; McGaw Park, IL 60085 USA; *** 2) Cat. 4341BSDF; Qty. 10; Safe-T Thoracentesis Tray; With Catheter and Sharps Injury Protection; Drug Free; CareFusion; 1500 Waukegan Road; McGaw Park, IL 60085 USA. Indicated for specimen retrieval and therapeutic drainage of the pleural cavity.
- Recall number
- Z-2193-2013
- Recalling firm
- Carefusion 2200 Inc
- Product
- *** 1) Cat. 4341B; Qty. 10; Thoracentesis Tray; With Catheter; Sterile; CareFusion; 1500 Waukegan Road; McGaw Park, IL 60085 USA; *** 2) Cat. 4341BSDF; Qty. 10; Safe-T Thoracentesis Tray; With Catheter and Sharps Injury Protection; Drug Free; CareFusion; 1500 Waukegan Road; McGaw Park, IL 60085 USA. Indicated for specimen retrieval and therapeutic drainage of the pleural cavity.
- Status
- Terminated
- Initiated
- 2013-08-28
- Posted
- 2013-09-11
- Terminated
- 2015-06-16
- Reason
- Catheters included in Thoracentesis Tray and Safe-T Thoracentesis Tray - Drug Free may occlude, preventing drainage of excessive fluid from the pleural cavity leading to a potential delay in the medical procedure and potentially serious patient complications.
- Root cause
- Process change control
- Action
- Recall notification letters dated 8/27/13 were distributed via courier to direct accounts. Consignees were/are requested to confirm receipt of the notification, examine their inventory for the recalled products and to destroy the products. After the products are destroyed, the consignees are requested to submit a completed response form to CareFusion for replacement with functional devices. Consignees who cannot destroy the affected products are requested to contact CareFusion for additional information.
- Quantity
- *** 1) 174,580 units; *** 2) 23,330 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MD, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV and Puerto Rico, and the country of Iceland.
- Lot, serial or model codes
- *** 1) Product Code: 4341B; Lots: 554361 (Expires 11/30/15); 550463 (Expires 11/30/15); 548675 (Expires 11/30/15); 545891 (Expires 10/31/15); 544746 (Expires 10/31/15); 543978 (Expires 10/31/15); 542297 (Expires 10/31/15); 541063 (Expires 10/31/15); 539859 (Expires 10/31/15); 539470 (Expires 10/31/15); 537723 (Expires 9/30/15); 536497 (Expires 9/30/15); 535600 (Expires 9/30/15); 534616 (Expires 9/30/15); 532917 (Expires 8/31/15); 531869 (Expires 8/31/15); 530729 (Expires 8/31/15); 528789 (Expires 8/31/15); 526724 (Expires 8/31/15); 525767 (Expires 8/31/15); 523275 (Expires 8/31/15); 522256 (Expires 8/31/15); 517247 (Expires 7/31/15); 514103 (Expires 7/31/15); 512953 (Expires 7/31/15); 509413 (Expires 5/31/15); 508928 (Expires 5/31/15); 505191 (Expires 5/31/15); 504328 (Expires 4/30/15); 499957 (Expires 4/30/15); 498168 (Expires 4/30/15); 495057 (Expires 3/31/15), 492933 (Expires 3/31/15); 487660 (Expires 2/28/15); 486920 (Expires 2/28/15); 483377 (Expires 11/30/14); 481000 (Expires 11/30/14); 478361 (Expires 11/30/14); 476883 (Expires 11/30/14); 474403 (Expires 10/31/14); 471330 (Expires 10/31/14); 469597 (Expires 10/31/14); 469561 (Expires 10/31/14); 467093 (Expires 10/31/14); 464366 (Expires 10/31/14); 463156 (Expires 8/31/14); 461311 (Expires 8/31/14); 458579 (Expires 8/31/14); 456606 (Expires 8/31/14); 453734 (Expires 8/31/14); 450541 (Expires 8/31/14); 450540 (Expires 8/31/14); 446817 (Expires 8/31/14); 444048 (Expires 8/31/14); 440805 (Expires 8/31/14); 439236 (Expires 8/31/14); 436820 (Expires 8/31/14) ;433808 (Expires 7/31/14); 429003 (Expires 7/31/14); 427075 (Expires 7/31/14); 424115 (Expires 7/31/14); 420673 (Expires 6/30/14); 419197 (Expires 6/30/14); 417436 (Expires 6/30/14); 413285 (Expires 6/30/14); 411301 (Expires 6/30/14); 408552 (Expires 7/31/14); 405227 (Expires 6/30/14); 403653 (Expires 6/30/14); 400855 (Expires 6/30/14); 399197 (Expires 6/30/14); 389066 (Expires 5/31/14); 387167 (Expires 5/31/14); 382554 (Expires 5/31/14); 380046 (Expires 4/30/14); 377477 (Expires 4/30/14); 375047 (Expires 4/30/14); 372900 (Expires 4/30/04); 370763 (Expires 4/30/14); 370204 (Expires 3/31/14); 368133 (Expires 3/31/14); 366017 (Expires 3/31/14); 363544 (Expires 3/31/14); 362824 (Expires 3/31/14); 362000 (Expires 2/28/14); 359063 (Expires 2/28/14); 357349 (Expires 2/28/14); 352522 (Expires 1/31/14); 350387 (Expires 1/31/14); 348043 (Expires 1/31/14); 347360 (Expires 1/31/14); 343651 (Expires 10/31/14); 341316 (Expires 1/31/14); 339958 (Expires 1/31/14); 335395 (Expires 12/31/13); 333550 (Expires 12/31/13); 330580 (Expires 11/30/13); 327592 (Expires 11/30/13); 326052 (Expires 10/31/13); 322688 (Expires 10/31/13); 320605 (Expires 10/31/13); 316289 (Expires 9/30/13); 313891 (Expires 9/30/13); 312365 (Expires 9/30/13); 308400 (Expires 9/30/13); L1B129 (Expires 8/31/13); L1B177 (Expires 8/31/13); L1B215 (Expires 10/31/13); L1B311 (Expires 10/31/13); L1C150 (Expires 12/31/13); L1C190 (Expires 12/31/13; L1C273 (Expires 8/31/13); L1C376 (Expires 8/31/13); *** 2) Product Code: 4341BSDF; Lots: 548682 (Expires 2/28/18); 542824 (Expires 2/28/18); 528896 (Expires 12/31/17); 523596 (Expires 12/31/17); 522262 (Expires 12/31/17); 511661 (Expires 10/31/17); 486156 (Expires 8/31/17); 485085 (Expires 8/31/17); 469770 (Expires 6/30/17); 456730 (Expires 5/31/17); 437220 (Expires 3/31/17); 435599 (Expires 3/31/17); 363941 (Expires 8/31/16) 362922 (Expires 8/31/16); 354615 (Expires 7/31/16); 349218 (Expires 7/31/16); 347700 (Expires 7/31/16); 347346 (Expires 6/30/16); L1B185 (Expires 2/28/16); L1B321 (Expires 3/31/16); L1C201 (Expires 3/31/16); L0A238 (Expires 1/13/15); L0C212 (Expires 3/31/15); L0D150 (Expires 4/30/15); L0E143 (Expires 5/31/13); L0J242 (Expires 7/31/15); L0J274 (Expires 8/31/15); L0K270 (Expires 9/30/15); L0L264 (Expires 10/31/15); L0P117 (Expires 11/30/15); L0P126 (Expires 11/30/15); L9A294 (Expires 1/31/14); L9C267 (Expires 3/31/14); L9D320 (Expires 4/30/14); L9E238 (Expires 5/31/14); L9H237 (Expires 6/30/14); L9J212 (Expires 7/31/14); L9K261 (Expires 9/30/14); L9P191 (Expires 12/31/14); L8K258 (Expires 8/31/13); L8L229 (Expires 9/30/13); L8S323 (Expires 12/31/13)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GCB
- Firm FEI
- 3008002452
- Firm city
- Vernon Hills
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28