Skip to the content

FDA Medical Device Recalls (openFDA)

Phasor Health, LLC: Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35

Recall number
Z-2192-2025
Recalling firm
Phasor Health, LLC
Product
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
Status
Open, Classified
Initiated
2025-03-21
Posted
2025-07-29
Terminated
not on file
Reason
Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
Root cause
Process change control
Action
Starting on 3/21/2025, device correction and removal began and a correction and removal notice was emailed. Customers were asked to do the following: 1) Review inventory (without opening the sterile transparent barrier) to determine if any of the affected lots remain in inventory. If yes, remove those lots from inventory. Do not use drills with reverse battery orientation. 2) If unable to review inventory soon, the Company will come to your site within 30 business days of this letter and inspect your inventory (within 20 minutes total) and replace any defective drills with new verified stock (same catalog number). 3) Contact firm at info@phasorhealth.com to report results of inventory inspection and to facilitate device return. 4) Complete and return the acknowledgement and receipt form via email to info@phasorhealth.com Please contact Firm at info@phasorhealth.com or (832) 982-1234 for more information or questions
Quantity
1,064
Distribution
US: PA, IN, AZ, MO, LA, FL, TX, MA, OH, TN, MI, WV, VA, CA, DE, NH, GA, SC, GU, NY, NJ, MN, WI, NE, WA, IL, AR, PR, VT
Lot, serial or model codes
UDI-DI: B775270S0, B775320L0, B7754500, B7755300, B7756350. Lot: 241002450, 241002530, 241009450, 241009530, 241016450, 241016530, 241023450, 241023530, 241030450, 241030530, 24112027S, 24121827S, 24122727S, 25012227S, 24082832L, 24102332L, 241227450, 241220635, 250122635, 241227530, 250122530.
510(k) numbers
["K161704"]
PMA numbers
not on file
Product code
HBE
Firm FEI
3016577622
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register