Abbott Diagnostics Technologies AS: Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.

Recall number
Z-2192-2020
Recalling firm
Abbott Diagnostics Technologies AS
Product
Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.
Status
Open, Classified
Initiated
2020-03-23
Posted
not on file
Terminated
not on file
Reason
The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.
Root cause
Device Design
Action
The recalling firm/manufacturer issued medical device correction letters dated 3/23/2020 on 3/23/2020 to their direct distributors requesting end user customer information. The recalling firm/manufacturer issued medical device correction letters to the end users dated 3/31/2020 via a third-party notification firm. The notification informed the end user of the issue and requested to be notified of the number of power supplies needed for replacement. The affected power supplies were to be discarded by the end user.
Quantity
42,237 instruments
Distribution
Distribution was made to NJ. There was no direct government/military distribution. Foreign distribution was made to Canada, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Denmark, Finland, France, Georgia, Germany, Guatemala, Hong Kong, Iceland, India, Italy, Ivory Coast, Japan, Korea, Madagascar, Malaysia, Mexico, Mongolia, Myanmar, The Netherlands, Norway, Oman, Pakistan, Peru, Philippines, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Timur, Trinidad & Tobago, United Kingdom, and Zimbabwe.
Lot, serial or model codes
All serial numbers are affected.
510(k) numbers
["K180296","K182988"]
PMA numbers
not on file
Product code
PDJ
Firm FEI
3003045237
Firm city
Oslo
Firm state
not on file
Firm country
Norway

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register