Sonosite, Inc.: SonoSite TEE, 8-3 MHz Transducer labeled: "Tee Transducer... SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021". The TEE/8-3 MHz Transducer is designed for 2D, M Mode, color Doppler, pulse wave (PW) Doppler and continuous wave (CW) Doppler by applying ultrasound energy through the esophagus or stomach of the patient into the heart. The TEE transducer is intended to be used on adults only. Backscattered ultrasound energy from the patient's heart is used to form images of the heart to detect abnormalities in structure or motion, to evaluate the velocity of blood flowing within the heart, and to obtain a color depiction of the velocities of blood flowing in the heart.

Recall number
Z-2191-2010
Recalling firm
Sonosite, Inc.
Product
SonoSite TEE, 8-3 MHz Transducer labeled: "Tee Transducer... SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021". The TEE/8-3 MHz Transducer is designed for 2D, M Mode, color Doppler, pulse wave (PW) Doppler and continuous wave (CW) Doppler by applying ultrasound energy through the esophagus or stomach of the patient into the heart. The TEE transducer is intended to be used on adults only. Backscattered ultrasound energy from the patient's heart is used to form images of the heart to detect abnormalities in structure or motion, to evaluate the velocity of blood flowing within the heart, and to obtain a color depiction of the velocities of blood flowing in the heart.
Status
Terminated
Initiated
2008-08-08
Posted
2010-08-09
Terminated
2011-08-24
Reason
The TEE transesophageal transducer, when connected to a SonoSite M-Turbo Ultrasound System, may exhibit erratic operation of the Scanplane Orientation Indicator on the system display. The indicator may move erratically and not accurately represent the orientation of the TEE transducer array.
Root cause
Other
Action
On August 21, 2008, SonoSite began calling 15 consignees who have purchased the TEE/8-3 MHz Transducer and an M-Turbo 1.1 system. The firm provided instruction to discontinue use the system to 11 consignees and left voice messages to the remaining four consignees. On August 21, 2008, SonoSite sent out the URGENT-MEDICAL DEVICE CORRECTION letter and the Urgent Field Safety Notice to all affected consignees in the US and in foreign countries. The URGENT-MEDICAL DEVICE CORRECTION letter describes the recall product as TEE/8-5 MHz Transducer, SonoSite Part Numbers P05183 and P06841. The letter informed the consignees that the TEE transesophageal transducers may exhibit erratic operation of the Scanplane Orientation Indicator on the system display when it connected to the SonoSite M-Turbo Ultrasound System. The letter instructs consignees to discontinue use of the TEE/8-3MHz Transducer with the M-Turbo (or Micromaxx system) immediately, while the firm is working to create a software solution to resolve the issue and offers customers a software upgrade within 6-8 weeks. The consignees are also advised to forward the recall notice to others who need to be aware of the issue or where the potentially affected devices have been distributed. Customers should call the Technical Support at 1-1-425-951-1330 or 1-877-657-8118 from Monday to Friday, 6 am-5 pm PST or e-mail to service@sonosite.com for any questions regarding the recall.
Quantity
357 units
Distribution
Worldwide distribution: USA and countries including Argentina, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Finland, France, Germany, India, Iran, Ireland, Israel, Italy, Korea, Netherlands, New Zealand, Poland, Slovenia, South Africa, Spain, Switzerland, and Thailand.
Lot, serial or model codes
Part number P05813 and P06841  Serial numbers: DOMESTIC: 03B3G4, 03B7BJ, 03BKK9, 03CWKJ, 037XL6, 034J4X, 034T2R, 038fdn, 03BKJH, 03B3FT, 039F3H, 03BC26, 03CJTD, 03CHVH, 039F3L, 034Z4M, 03CWKD, 0354HB, 039F3J, 039F3Q, 03B3G3, 0354HC, 03B7B3, 0354HD, 03BKJ0, 03BKJK, 0356LC, 0360X5, 03BC2B, 03CVVH, 03CCBT, 0377LT, 038WXY, 03BC25, 037MR9, 037XPX, 03B3FY, 037XQ1, 03B3FX, 0392RC, 03B3G0, 0356FB, 035G72, 035NT0, 038WTJ, 03BKJF, 038WY1, 038WY3, 038WY5, 03CHVG, 03B3FW, 03CJ7D, 0392R8, 03CHVM, 03B3FV, 03936X, 0361FF, 0354J0, 0377LQ, 03CWCX, 0379T3, 037F5D, 03BC2G, 037X7R, 035NT3, 0384GM, 0392LP, 0360X7, 0376PB, 03936W, and 03BKKC.    INTERNATIONAL: 035G71, 038FDP, 039V7L, 0356C3, 037XQ5, 0384K6, 0392RH, 03CJ83, 0377LN, 0392RG, 039KD3, 039V7P, 039V7T, 03BC24, 03BC2D, 03BKJJ, 03BKKY, 038WRH, 03CCBQ, 0384FH, 039KD4, 03BC20, 034PGZ, 034PH0, 03CCBR, 03BC23, 03CWKL, 034T2T , 03B7B2, 0384FL, 0384GN, 0384K4, 0384K5, 038WRD, 0390MH, 039190, 03918Z, 03BKKW, 03CJ82, 03CJ7B, 03CJ81, 03CJ86, 03CJ87, 03CJ88, 037XL7, 0384BX, 038WY7, 034PGQ, 0392RL, 034PGV, 03BC2C, 0392R9 , 039196 ,03CHVF, 03CJTB, 03CJTC, 034T30, 03570P, 0354HF, 0377LP, 038FL4 , 03CCBP, 03CWKF, 037MR7, 037XPW, 037MR6, 0384FJ, 034TN9, 0356HM, 038WTM, 0356GL, 034T32 , 039198, 0360YT , 0384CZ, 0384D1, 0361G5, 03B7B2, 03BKJG, 03CHVD, 03CJ7C, 03CJ84, 03CJ8B, 03CVVJ, 034T2W , 0356BV , 0377LV, 037MRC, 038WRJ, 039F3R, 0361DN, 036GYG, 037LTV, 037MR8, 037XQ3, 034PGX 035TZK, 034P7Y, 034PGM, 034PGP, 0356BW, 035G9X, 035NT2, 0361G6, 036GYD, 037F1Y, 037LTQ, 037LTT, 037LTY, 037M8T, 037XPY, 038WY2, 039V7M, 03CWD1, 03CWKM, 038XKX, 034NMJ, 0356C2, 034T2V, 034T2X, 034TNW, 034TNZ, 0354H9, 0356LD, 0360YV, 037XL8, and 0390MC.
510(k) numbers
["K071134"]
PMA numbers
not on file
Product code
IYO
Firm FEI
3002087856
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register