FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214P; b) ABLATION PACK, Model Number DYNJ82725

Recall number
Z-2189-2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214P; b) ABLATION PACK, Model Number DYNJ82725
Status
Open, Classified
Initiated
2023-05-18
Posted
2023-07-20
Terminated
not on file
Reason
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Root cause
Device Design
Action
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Quantity
316 units
Distribution
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Lot, serial or model codes
all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Number CDS840214P, UDI/DI (EA) 10193489443196, UDI/DI (CS) 40193489443197; b) Model Number DYNJ82725, UDI/DI (EA) 10195327207380, UDI/DI (CS) 40195327207381
510(k) numbers
not on file
PMA numbers
not on file
Product code
OKV
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register