Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System

Recall number
Z-2188-2016
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System
Status
Terminated
Initiated
2016-02-24
Posted
2016-07-18
Terminated
2017-04-25
Reason
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
Root cause
Radiation Control for Health and Safety Act
Action
Toshiba Medical Systems (TAMS) planned action to bring product into Compliance: 1.TAMS will install the software and perform testing to ensure the software update was effective. 2. The customer notification letter which includes a statement that TAMS will, without charge, remedy the defect or bring the product into compliance. . If you have any questions regarding this, please feel free to contact your local Toshiba representative at (800) 521-1968.
Quantity
23
Distribution
US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
Lot, serial or model codes
Model Number - XIDF-AWS801
510(k) numbers
["K150967"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register