Skip to the content

FDA Medical Device Recalls (openFDA)

Wright Medical Technology Inc: CLAW HEX SCREW CHARLOTTE " F&A SYSTEM CONTENTS: 1 EACH Foot and Ankle Screw ... IMPLANT MATERIAL: Stainless Steel Ref: 4013-3520 ... Size 3.5MM, Length 20 MM, Use with Clava 3.5MM - Non-Sterile. Implantation: used in Food and Ankle surgeries.

Recall number
Z-2188-2010
Recalling firm
Wright Medical Technology Inc
Product
CLAW HEX SCREW CHARLOTTE " F&A SYSTEM CONTENTS: 1 EACH Foot and Ankle Screw ... IMPLANT MATERIAL: Stainless Steel Ref: 4013-3520 ... Size 3.5MM, Length 20 MM, Use with Clava 3.5MM - Non-Sterile. Implantation: used in Food and Ankle surgeries.
Status
Terminated
Initiated
2010-06-07
Posted
2010-08-09
Terminated
2011-11-23
Reason
One lot of CLAW Hex Foot and Ankle Screws was incorrectly packaged containing Multi-use Compression Screws.
Root cause
Packaging process control
Action
The firm initiated their recall to their sales representatives (distributors) via telephone on 06/07/2010 and followed with a letter dated June 15, 2010 sent via Fed Ex on 06/28/2010. The hospital involved was also notified by letter on 06/29/2010. The letter described the product, the problem,and the action to be taken by the customers. The customers were instructed to immediately return any affected product to Wright as soon as possible, and to complete an attached response form confirming receipt of the notice and return a copy by fax to 901-867-7401. Customers may contact customer service for return instructions and replacement inventory at 800-238-7117. If you have any questions regarding this matter, contact Debby Daurer at 800-874-5630.
Quantity
8 units
Distribution
Nationwide Distribution in States of FL, PA, KS, IN, MO, and CA
Lot, serial or model codes
Item # 40133520 Lot # 0501148499
510(k) numbers
["K080295"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1043534
Firm city
Arlington
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register