Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. These screws are intended for bicortical anchorage.

Recall number
Z-2187-2012
Recalling firm
Zimmer, Inc.
Product
The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. These screws are intended for bicortical anchorage.
Status
Terminated
Initiated
2012-07-13
Posted
2012-08-10
Terminated
2013-09-24
Reason
Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
Root cause
Labeling mix-ups
Action
The firm, Zimmer, sent an "URGENT MEDICAL DEVICE RECALL" letter dated July 13, 2012, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to locate the implicated lot in their inventory and user accounts; return all packaged units of this lot; and provide Zimmer with the name, complete mailing address, and phone number of each user facility and surgeon that have used the product via E-mail to dawn.kindle@zimmer.com. Zimmer will send a follow-up Urgent Recall notification to these sub-accounts with instructions to verify the screw length of any screws already in the implant rack inventory prior to use in surgery. All implicated product is to be returned. The customers were also instructed to complete and return the Inventory Return Certification form to Zimmer via fax at +1-574-372-4265. If you have any questions about this correction and/or product removal, contact Zimmer at +1-877-946-2761.
Quantity
24 to US Distributor with 7 further distributed
Distribution
Nationwide distribution: USA including states of: AZ, IL, IN, ME and NJ.
Lot, serial or model codes
Part number 02.03155.026, lot 2649941
510(k) numbers
["K042695"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28