AGFA Corp.: IMPAX Cardiovascular (CV) Reporting Reporting tool used for structured reporting and structured report outputs (including XML style sheet customization and PDF)

Recall number
Z-2184-2012
Recalling firm
AGFA Corp.
Product
IMPAX Cardiovascular (CV) Reporting Reporting tool used for structured reporting and structured report outputs (including XML style sheet customization and PDF)
Status
Terminated
Initiated
2012-07-27
Posted
2012-08-10
Terminated
2015-01-13
Reason
Content entered into the "Conclusions" free text box on the Report Writer screen was not represented on the final printed report.
Root cause
Software change control
Action
AGFA Healthcare sent an URGENT FIELD SAFETY NOTICE dated July 27, 2012, to all affected customers. The letter described the safety alert and mitigation. Customer were asked to complete the Acknowledgment form and fax back to 864-421-1664 that the information was received and understood. For questions customers were instructed tol call 401-604-2180. For questions regarding this recall call 864-421-1754.
Quantity
280
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Lot, serial or model codes
Model # L9M2100 - All Software versions of IMPAX CV Reporting
510(k) numbers
["K050228"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register