Skip to the content

FDA Medical Device Recalls (openFDA)

KCI USA, Inc.: KCI RotoProne Critical Care Therapy System buckles for securing patient to the surface during proning therapy, Model 209500. Indicated for treatment and prevention of pulmonary complications.

Recall number
Z-2182-2011
Recalling firm
KCI USA, Inc.
Product
KCI RotoProne Critical Care Therapy System buckles for securing patient to the surface during proning therapy, Model 209500. Indicated for treatment and prevention of pulmonary complications.
Status
Terminated
Initiated
2011-02-18
Posted
2011-05-10
Terminated
2011-10-24
Reason
Buckles used to secure patient to surface during proning therapy can come open if not properly engaged by user.
Root cause
Device Design
Action
Firm sent a letter with the "Warning: Improper Insertion of the RotoProne Therapy System Buckle Can Lead To Unintentional Bed Exit" to all consignees on 2/23/2011. The letter identified the affected product and the reason for the recall. Customers were instructed to read the attached product warning regarding the proper method to attach the RotoProne buckles and to post it with the device to ensure continued proper use of the buckles. If customers had questions regarding the proper use of the buckles, they were to contact KCI Customer Technical Service at 1-800-275-4524.
Quantity
229 units
Distribution
Worldwide Distribution - USA and Kuwait.
Lot, serial or model codes
Serial numbers: RRFK0036-RRFK00047; RRFK00049-RRFK00115; RRFK00117-RRFK00128; RRFK00141-RRFK00211; RRFK00213-RRFK00233; RRFK00235; RRFK00237-RRRFK00254; RRFK00256-RRFK00261; RRFK00263-RRFK00267; RRFK00269-RRFK00270; RRFK00273-RRFK00280; RRFK00282-RRFK00284; RRFK00287-RRFK00290; and RRJK00002.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IKZ
Firm FEI
1625774
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register