Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Neurovascular: Target Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Recall number
Z-2181-2016
Recalling firm
Stryker Neurovascular
Product
Target Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Status
Terminated
Initiated
2016-06-06
Posted
2016-07-11
Terminated
2017-03-03
Reason
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Root cause
Process change control
Action
Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instructed to complete the attached customer response form and return the completed form to their nominated Stryker Representative.
Quantity
45,302 (8,188 in the U.S.) - total, all model numbers
Distribution
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Lot, serial or model codes
MODEL Number:M0035452060;  Lot Number/Expiration Date: 18777380 31-Dec-18 18538894 31-Aug-18 18726354 30-Nov-18 18830541 31-Jan-19 18802659 31-Jan-19 18776332 31-Dec-18 18499021 31-Jul-18 18743910 30-Nov-18 18725944 30-Nov-18 18454649 30-Jun-18 18794568 31-Jan-19 18538832 31-Aug-18 18829512 31-Jan-19 18801344 31-Jan-19 18709049 30-Nov-18 18693095 30-Nov-18 18499207 31-Jul-18 19032899 30-Apr-19 19047487 31-May-19 18852469 28-Feb-19 18954631 31-Mar-19 18906669 28-Feb-19 18877336 28-Feb-19 19037939 31-May-19 18973118 30-Apr-19 18899998 28-Feb-19 18900840 28-Feb-19 19046456 31-May-19 18972772 30-Apr-19 18955537 31-Mar-19 19005724 30-Apr-19 18935230 31-Mar-19 18925282 31-Mar-19 18876468 28-Feb-19 19032992 30-Apr-19.
510(k) numbers
["K113412","K153658"]
PMA numbers
not on file
Product code
HCG
Firm FEI
3008853977
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register