Skip to the content

FDA Medical Device Recalls (openFDA)

Vascular Solutions, Inc.: Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.

Recall number
Z-2180-2017
Recalling firm
Vascular Solutions, Inc.
Product
Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.
Status
Terminated
Initiated
2017-04-12
Posted
not on file
Terminated
2018-04-11
Reason
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Root cause
Packaging
Action
Customers were sent a Vascular Solutions "Urgent Medical Device Removal letter" dated April 12, 2017. The letter described the problem and the product involved in the recall. Advised consignees to remove and secure all affected product and to complete and return the "VSI Account Inventory Form". For questions contact your local Vascular Solutions Sales Representative. For further questions, please call (763) 656-4300.
Quantity
4,110
Distribution
US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
Lot, serial or model codes
Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261.
510(k) numbers
["K042449"]
PMA numbers
not on file
Product code
MAV
Firm FEI
3002827704
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register