FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT Oncologist software applications used with SIEMENS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
- Recall number
- Z-2180-2014
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT Oncologist software applications used with SIEMENS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
- Status
- Terminated
- Initiated
- 2014-06-17
- Posted
- 2014-08-10
- Terminated
- 2014-12-15
- Reason
- Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the inability to unload patient data.
- Root cause
- Software design
- Action
- In June, 2014, an urgent medical device correction field safety notice of improvement was sent to end users informing them of the product, problem, and action to be taken to avoid the potential safety issue.
- Quantity
- 56
- Distribution
- UT, WI, NY
- Lot, serial or model codes
- Syngo RT Therapist (RTT 4.3.SP1) - model no. 08162815 Syngo RT Oncologist (RTO 4.3.SP1) - model no. 10568604 Syngo RT Therapist (RTT 4.3.SP1) for U.S. affected units only: 10010 10046 10586 No Syngo RT Oncologist (RTO 4.3.SP1) listed for U.S.
- 510(k) numbers
- ["K072485"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28