Skip to the content

FDA Medical Device Recalls (openFDA)

Digisonics, Inc: Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a digital image management and reporting system with an associated clinical database. It provides capture, review, and storage of images together with report creation and data basing.

Recall number
Z-2180-2011
Recalling firm
Digisonics, Inc
Product
Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a digital image management and reporting system with an associated clinical database. It provides capture, review, and storage of images together with report creation and data basing.
Status
Terminated
Initiated
2009-01-20
Posted
2011-05-10
Terminated
2011-05-31
Reason
Data from one patient's study had the potential for being put into another patient's report.
Root cause
Device Design
Action
Digisonics notified end users via e-mail January 20, 2009 and January 21, 2009, that Data from one patient's study was being put into another patient's report. The problem occurred when a System bug was found in DigiView Release 3.6.4.4, and higher versions in which under certain circumstances (specific workflow) a failure to switch to a patient's images occurs. Two software patch versions will be made available to sites on or after February 5, 2009, to correct the reported problem. If you have any further questions, please contact Digisonics Customer Support at (800 ) 940-3240.
Quantity
70 units
Distribution
Worldwide Distribution -- USA including States of : OH, NC, LA, TX, CA, MO, NY, MD, MA, DC, IA, AZ, PA, MI, NC, FL, IN, SC, ID, MN, AR, OK, IL, CT, and countries of Canada, Lebanon, and New Zealand.
Lot, serial or model codes
DigiView Rel. 3.6.4.4, 3.6.4.6, and 3.6.5.2.
510(k) numbers
K970402
PMA numbers
not on file
Product code
LLZ
Firm FEI
1626313
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source