FDA Medical Device Recalls (openFDA)
CareFusion 303, Inc.: BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
- Recall number
- Z-2179-2025
- Recalling firm
- CareFusion 303, Inc.
- Product
- BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
- Status
- Open, Classified
- Initiated
- 2025-07-08
- Posted
- 2025-08-06
- Terminated
- not on file
- Reason
- Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Root cause
- Under Investigation by firm
- Action
- On December 12, 2025, CareFusion 303, Inc. (BD) issued a updated "Urgent Medical device Recall Notice via E-Mail and physical mail. The updated notice includes important information on infusion sets as well as labeling updates. Previous notifications were sent on July 8,2025, July 17, 2025, September 11, 2025 and September 12, 2025. CareFusion ask consignees to take the following actions: 1. Review your inventory for the catalog numbers included in the updated letter. Remove and discard any remaining inventory per your facility guidelines. 2.To receive credit or replacement product, please return the completed customer response form attached to this notice. 3.Please contact your BD Clinical Consultants and Account Executives for alternative sets information. 4. Review the updated sets performance data included in the letter. 5. Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 6. Distributors should provide a copy of this notice to all customers who may have purchased an Alaris Pump Module. If you require further assistance, please contact: BD Technical Support at Phone: 888-812-3229; Phone hours: 6:00am to 5:00pm PT Monday Friday and Email: DL-US-INF-TechSupport@bd.com.
- Quantity
- 33,162,680 units total
- Distribution
- US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
- Lot, serial or model codes
- BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046
- 510(k) numbers
- ["K211218"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 2016493
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28