Datascope Corporation: Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 4906 PEDIARIC CPB PACK, Part Number 701056041 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only and is not to be re-sterilized by the user. The product consists of tubing and connectors with connected oxygenators, reservoirs, filters and other cardiopulmonary bypass components assembled into user specified circuits for procedures requiring extracorporeal support for up to six hours. See Attachment 6 for a copy of MCP Instructions for Use.

Recall number
Z-2179-2018
Recalling firm
Datascope Corporation
Product
Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 4906 PEDIARIC CPB PACK, Part Number 701056041 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only and is not to be re-sterilized by the user. The product consists of tubing and connectors with connected oxygenators, reservoirs, filters and other cardiopulmonary bypass components assembled into user specified circuits for procedures requiring extracorporeal support for up to six hours. See Attachment 6 for a copy of MCP Instructions for Use.
Status
Terminated
Initiated
2017-07-27
Posted
not on file
Terminated
2020-02-07
Reason
There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
Root cause
Nonconforming Material/Component
Action
On 7/27/17, an Urgent Safety Alert was issued to customers of the Cardiovascular Procedure Kit with LN130B Over-pressure Safety Valves. The safety alert instructed customers to perform the following: Please examine your inventory immediately to determine if you have unused kits of the batch numbers listed. If so, please follow the recommended actions outlined in the attached Terumo Cardiovascular Systems Urgent Safety Alert. It is possible that custom tubing kits with the affected valves may have already been used without issue. Customers were instructed to test the valve prior to use. The notice was sent to customers again on 10/3/17 and 12/12/17. On 1/11/18, Urgent Medical Device Recall letters were sent to customers. The letters included the following instructions: Terumo CVS now recommends removal of any remaining product from your inventory. A review of our records indicates that you have received the following Maquet/Getinge Custom Tubing Kit batch numbers affected by the Terumo Cardiovascular Systems Medical Device Recall (refer to information provided in the table below). Please examine your inventory immediately to determine if you have unused custom tubing kits of the batch numbers listed. If so, please remove the affected products, quarantine them and follow the instructions for returning the product to Maquet/Getinge for full credit. It is possible that custom tubing kits with the affected valves may have already been used without issue. The Terumo Cardiovascular Systems Medical Device Recall affects the Custom Tubing Kit with LN130B Over-pressure Safety Valve distributed by Maquet/Getinge having lot/batch numbers listed above; no other product lot/batch numbers that were delivered to you are affected. Instructions for returning affected product: 1. Review this notice and Terumos Medical Device Recall Notice (attached). 2. Ensure that all users (i.e. perfusionist, operating room staff and biomedical staff) receive notice
Quantity
27
Distribution
The products were distributed to the following US states: IA and FL.
Lot, serial or model codes
UDI 0607567203600
510(k) numbers
["K080592"]
PMA numbers
not on file
Product code
DWE
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register