Medtronic MiniMed Inc.: Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.

Recall number
Z-2179-2014
Recalling firm
Medtronic MiniMed Inc.
Product
Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.
Status
Terminated
Initiated
2014-07-21
Posted
2014-08-10
Terminated
2014-10-08
Reason
Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.
Root cause
Device Design
Action
Medical Device Recall letters dated July 21, 2014 andJuly 22, 2014 were sent to distributors, patients, and healthcare providers who purchased the Guardian REAL time monitors. The letters inform the customers of the problems identified and the actions to be taken. Distributors with questions are instructed to email rs.nrdlrinbox@medtronic.com. If healthcare professionals want to speak to Medtronic's medical officers then they are instructed to call (818) 576-4211, ext. 64511. Customers and healthcare professionals' patients are instructed to call Guardian Recall Hotline at (800) 646-4633, ext. 21090, 8am-6pm, CST.
Quantity
86 units
Distribution
Worldwide:Distribution - US Nationwide and the countries of Croatia, Denmark, Finland, Germany, Netherlands, Norway, Sweden, Switzerland.
Lot, serial or model codes
CSS7100 and CSS7100K
510(k) numbers
not on file
PMA numbers
["P980022"]
Product code
MDS
Firm FEI
3003166194
Firm city
Northridge
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register