Toshiba American Medical Systems Inc: Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System

Recall number
Z-2179-2011
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System
Status
Terminated
Initiated
2010-04-20
Posted
2011-05-10
Terminated
2011-05-25
Reason
There is a possibility of the system shutting down owing to the power supply unit causing the protection function to activate thus turning off the system prematurely.
Root cause
Device Design
Action
The recall was initiated on 4/20/10 with Toshiba America Medical Systems (TAMS) forwarding a Medical Device Correction with a Customer Reply Form (via US Postal Service for US Customers) to all customers who purchased the SSH-880CV Aplio Artida Diagnostic Ultrasound System. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to avoid this problem until corrective action is taken, in case this problem occurs; please contact their Toshiba Representative. If emergency use is necessary, use the following procedure which allows the system to be used temporarily: 1. Turn OFF the breaker at the back of the system and confirm that the error code near the breaker disappears. 2. Turn ON the breaker again to supply power to the system. Customers were instructed to complete and return the attached form and fax it to the toll free number at the top of the form. This form can also be sent via email to raffairs@tams.com. If customers had any questions they can call Paul Biggins at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Quantity
14
Distribution
USA, including Puerto Rico.
Lot, serial or model codes
W1F09X2272, W1F0972220, W1F0992260, W1F0992262, W1H1012311, W1F0972219, W1F09Y2281, W1F0972218, W1F09X2270, W1F09X2271, W1F09Y2280, W1F09X2268, W1F09X2273, W1F09Y2282, W1F09Z2296, W1F09Z2297, W1F0992261
510(k) numbers
["K090158"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register