FDA Medical Device Recalls (openFDA)
Toshiba American Medical Systems Inc: Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System
- Recall number
- Z-2179-2011
- Recalling firm
- Toshiba American Medical Systems Inc
- Product
- Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System
- Status
- Terminated
- Initiated
- 2010-04-20
- Posted
- 2011-05-10
- Terminated
- 2011-05-25
- Reason
- There is a possibility of the system shutting down owing to the power supply unit causing the protection function to activate thus turning off the system prematurely.
- Root cause
- Device Design
- Action
- The recall was initiated on 4/20/10 with Toshiba America Medical Systems (TAMS) forwarding a Medical Device Correction with a Customer Reply Form (via US Postal Service for US Customers) to all customers who purchased the SSH-880CV Aplio Artida Diagnostic Ultrasound System. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to avoid this problem until corrective action is taken, in case this problem occurs; please contact their Toshiba Representative. If emergency use is necessary, use the following procedure which allows the system to be used temporarily: 1. Turn OFF the breaker at the back of the system and confirm that the error code near the breaker disappears. 2. Turn ON the breaker again to supply power to the system. Customers were instructed to complete and return the attached form and fax it to the toll free number at the top of the form. This form can also be sent via email to raffairs@tams.com. If customers had any questions they can call Paul Biggins at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
- Quantity
- 14
- Distribution
- USA, including Puerto Rico.
- Lot, serial or model codes
- W1F09X2272, W1F0972220, W1F0992260, W1F0992262, W1H1012311, W1F0972219, W1F09Y2281, W1F0972218, W1F09X2270, W1F09X2271, W1F09Y2280, W1F09X2268, W1F09X2273, W1F09Y2282, W1F09Z2296, W1F09Z2297, W1F0992261
- 510(k) numbers
- ["K090158"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28