Covidien LLC: Monoject 0.9% Sodium Chloride Flush Syringe, 12 mL Syringe with 10 mL Fill Product ID: 8881570121 Product Usage: Both 0.9% Sodium Chloride Flush syringes and 10U/mL or 100U/mL Heparin Lock Flush syringes are sterile, single-use products which are intended for use in flushing and locking compatible intravenous administration sets and indwelling intravenous access devices
- Recall number
- Z-2178-2013
- Recalling firm
- Covidien LLC
- Product
- Monoject 0.9% Sodium Chloride Flush Syringe, 12 mL Syringe with 10 mL Fill Product ID: 8881570121 Product Usage: Both 0.9% Sodium Chloride Flush syringes and 10U/mL or 100U/mL Heparin Lock Flush syringes are sterile, single-use products which are intended for use in flushing and locking compatible intravenous administration sets and indwelling intravenous access devices
- Status
- Terminated
- Initiated
- 2013-08-16
- Posted
- 2013-09-10
- Terminated
- 2015-10-21
- Reason
- Monoject prefill flush syringes may contain non-sterile water and labeled as saline or heparin
- Root cause
- Process control
- Action
- Covidien sent an Urgent Medical Device Recall letter dated August 16, 2013 via Federal Express to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter requested that customers immediately quarantine and discontinue use of the affected products, return any affected product by following the instructions provided in the letter, and complete and return the Recalled Product Return Form to Covidien. The letter states that if the affected products have been distributed, assure that all customers are notified and affected product is removed from the distribution channel as well as from customer stock. Questions or concerns, please do not hesitate to contact your Covidien representative or Covidien Customer Service, Monday through Friday, 8am- 6:30pm ET, at (800) 962-9888, option 1, and then option 2. Firm issued Press on 8/19/13
- Quantity
- 3,235,860 units
- Distribution
- Worldwide Distribution - USA Nationwide and the country of Bermuda
- Lot, serial or model codes
- Lot Numbers: 13A0084N 13A0094 13B0364 13C0514
- 510(k) numbers
- ["K032438"]
- PMA numbers
- not on file
- Product code
- NGT
- Firm FEI
- 1282497
- Firm city
- Mansfield
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28