Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems

Recall number
Z-2177-2025
Recalling firm
Steris Corporation
Product
Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
Status
Open, Classified
Initiated
2025-06-23
Posted
2025-07-25
Terminated
not on file
Reason
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Root cause
Process control
Action
Steris issued an URGENT MEDICAL DEVICE RECALL NOTICE dated 06/17/2025 to its consignees on 06/23/2025 disseminated via FedEx. The notice explained the issue, potential risk, and requested the following: User Action: Please ensure the following steps are completed: 1. Review the revised Preventive Maintenance Manual and Interval Based Check List language depending on which HarmonyAIR Surgical Lighting System model(s) are in use at your facility (A-Series and/or E-Series) (please reference Attachments B-E of the letter for maintenance instructions). 2. Immediately complete the maintenance activities as instructed, and then again annually going forward. 3. Complete the Medical Device Recall Response Form included with this letter. 4. Return the completed Response Form via email to: Regulatory_Compliance@STERIS.com, or via fax to 440-392-8963. 5. Distributors: If you have further distributed this product, please identify your Customer(s), and notify them of this recall.
Quantity
167 units
Distribution
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Lot, serial or model codes
UDI: 00724995203245: Serial Numbers: 0400224028, 0401625167, 0401625171, 0401625180, 0401625185, 0401625190, 0402022165, 0402022169, 0402422003, 0402622007, 0402925148, 0403122053, 0403122057, 0403125009, 0403223079, 0403223083, 0403223087, 0403223091, 0403321066,0403321070, 0403423063, 0403423067, 0403423071, 0403423075, 0403423079, 0403423083, 0403423087, 0404022003, 0404622011, 0404622014, 0404622017, 0404622020, 0404622023, 0404822104, 0405025032, 0405924100, 0405924103, 0406624134, 0406624138, 0406624142, 0406722052, 0406722056, 0406722060, 0406722064, 0406823005, 0407625022, 0408025008, 0408222120, 0408222124, 0408222128, 0408222132, 0408922093, 0408922097, 0408922101, 0409622008, 0409822014, 0409822018, 0410822042, 0410822046, 0410823068, 0410923149, 0410923152, 0410923155, 0413223057, 0413223060, 0414223170, 0414223174, 0414223178, 0414223182, 0414223186, 0414223190, 0414223194, 0414223198, 0414323003, 0414323006, 0415223065, 0415223068, 0415223071, 0415223074, 0415923036, 0416924069, 0416924074, 0417422053, 0418023091, 0418023095, 0418023099, 0418023103, 0418023107, 0418023111, 0419223082, 0419523010, 0419724177, 0419924007, 0420223048, 0420223051, 0420223054, 0420223057, 0420223060, 0420823170, 0421322041, 0421322045, 0421623206, 0421623211, 0421623216, 0421623221, 0421623226, 0421623231, 0421623236, 0421623241, 0422923010, 0423023059, 0423023304, 0423023307, 0423023379, 0423023384, 0423023389, 0423023394, 0423023399, 0423023451, 0423422042, 0423422045, 0423422048, 0423924004, 0424224127, 0424224131, 0424323096, 0426223004, 0426223008, 0427922038, 0427922042, 0428824102, 0428924075, 0428924079, 0428924083, 0429823019, 0429823023, 0430323199, 0430523194, 0430523197, 0430722034, 0431123070, 0432324031, 0432324156, 0432324159, 0432324162, 0432324165, 0432722064, 0433724005, 0433724010, 0433724015, 0434023084, 0434023087, 0434023090, 0434023111, 0434023116, 0434023121, 0434622027, 0434622031, 0434622035, 0434722013, 0434722159, 0434722163, 0436223068, 0436624179, 0436624184, 0436624189, 0436624194;
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSY
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register