FDA Medical Device Recalls (openFDA)
AGFA Healthcare Corp.: DX-D 600 - DXD Imaging Package Product Usage:Is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities, check abdomen and other both parts.
- Recall number
- Z-2175-2014
- Recalling firm
- AGFA Healthcare Corp.
- Product
- DX-D 600 - DXD Imaging Package Product Usage:Is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities, check abdomen and other both parts.
- Status
- Terminated
- Initiated
- 2014-07-02
- Posted
- 2014-08-10
- Terminated
- 2015-06-04
- Reason
- Some customers reported that while using the DX-D600 in room or exam preparation, the Overhead Tube Crane (OTC) showed increased velocity and uncontrolled longitudinal movement.
- Root cause
- Device Design
- Action
- AGFA HealthCare sent an URGENT FIELD SAFETY NOTICE letter dated July 2, 2014 to affected customers. The letter described the affected product, problem and actions to be taken. Customers were instructed to complete and return the Acknowledgment form, via FAX-Back or email. A Mandatory Service Bulletin DIS012.14E/No. 32, was released June 30, 2014, to implement the mandatory upgrade of full automatic DX-D600 systems to version 3.6 to prevent unexpected system movements. This Mandatory Service Bulletin was issued to Agfa DR Specialists to perform the service. A 2nd release will be implemented to upgrade semi-automatic systems. For questions contact your local AGFA HealthCare organization 1-877-777-2432.
- Quantity
- 17 units
- Distribution
- Worldwide Distribution - US Natinwide in the state of GA, IN, KS, KY, MA, NJ, SC, VA, DC and the country of Canada
- Lot, serial or model codes
- Serial Nos. A5430000089, A5430000131, A5430000015, A5430000095, A5430000052, A5430000061, A5430000079, A5430000030, A5430000042, A5430000066, A5430000081, A5430000085, A5430000096, A5430000020, A5430000067, A5430000125, A5430000105
- 510(k) numbers
- ["K112670"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as