FDA Medical Device Recalls (openFDA)
Philips Healthcare Inc.: Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information Portfolio Release D.03; 865209 IntelliVue Clinical Information Portfolio Release E.0 and E.01; 865300 IntelliVue Clinical Information Portfolio NICU Link
- Recall number
- Z-2175-2011
- Recalling firm
- Philips Healthcare Inc.
- Product
- Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information Portfolio Release D.03; 865209 IntelliVue Clinical Information Portfolio Release E.0 and E.01; 865300 IntelliVue Clinical Information Portfolio NICU Link
- Status
- Terminated
- Initiated
- 2011-03-16
- Posted
- 2011-05-10
- Terminated
- 2020-08-13
- Reason
- Edits to an order can result in unintended scheduled interventions and a patient may receive additional interventions or medication doses that are not intended.
- Root cause
- Software design
- Action
- Philips Healthcare notified accounts on 3/16/11 via the Urgent Medical Device Correction Notification. Customer notifications in the United States will be delivered via UPS. Customer notifications outside the US will be managed by Philips representatives in each affected geography. Customers will be asked to follow the guidelines in the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. All customers potentially affected by this issue will receive the software correction as part of the field action.
- Quantity
- 387 units
- Distribution
- Worldwide distribution: USA, Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Kenya, Latvia, Malaysia, Netherlands, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
- Lot, serial or model codes
- versions D.03, E.0, or E.01 of the software
- 510(k) numbers
- ["K100272","K992636"]
- PMA numbers
- not on file
- Product code
- NSX
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28