Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Model 8731SC Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.

Recall number
Z-2174-2008
Recalling firm
Medtronic Neuromodulation
Product
Model 8731SC Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.
Status
Terminated
Initiated
2008-04-14
Posted
2008-08-27
Terminated
2008-09-10
Reason
The SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Root cause
Device Design
Action
Medtronic informed all Belgium implanting and managing physicians of the SynchroMed II infusion system. Each customer was mailed a Dear Healthcare Professional letter (Medtronic Urgent: Field Safety Notice) on 4/3/08 providing important safety information and patient management recommendations regarding the device incompatibility. The firm recommends that healthcare professionals do not use SC catheters if a connection with an implantable access port is required. Also, if a continuous infusion trial is required with use of an implantable access port, they should use a model 8711 intrathecal catheter. The firm states that for patients who are currently undergoing a continuous trial with an implantable port, replacing the catheter connector before connecting to the SynchroMedd II using the Medtronic model 8578 revision kit is recommended. Medtronic also recommends discussing the risks of connection failure with patients. Local field reps will personally follow-up with affected customers to confirm and document notification.
Quantity
232 total from all models
Distribution
International Distribution --- country of Belgium.
Lot, serial or model codes
All lots numbers included
510(k) numbers
not on file
PMA numbers
["P860004"]
Product code
LKK
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register