FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp. (NeuroSciences): Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane catheter for diverting cerebrospinal fluid (CSF) from the ventricles of the brain through a series of drainage holes near the catheter's bullet shaped tip.
- Recall number
- Z-2173-2026
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences)
- Product
- Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane catheter for diverting cerebrospinal fluid (CSF) from the ventricles of the brain through a series of drainage holes near the catheter's bullet shaped tip.
- Status
- Open, Classified
- Initiated
- 2026-04-10
- Posted
- 2026-05-13
- Terminated
- not on file
- Reason
- Out-of-specification endotoxin result that did not meet the acceptance criteria.
- Root cause
- Process control
- Action
- Integra notified consignees on about 04/10/2026 via letter. Consignees at medical facilities were instructed to examine inventory for affected units, remove them immediately from service and quarantine them, forward the notification to those who utilize the product so they are aware of the recall and can identify any affected product that may remain in clinical areas, and complete and return the "Medical Facility Acknowledgement Form provided. Integra Customer Care will contact consignees who have noted affected product on hand and provide a Return Material Authorization (RMA) number and directions to return the affected product. Distributors were instructed to remove the product from further distribution, complete and return the Acknowledgement Form, check customer traceability records and notify them if they have any shipments of affected product, modify the acknowledgement form to provide to their customers, collect completed response forms and affected product from customers and report to Integra on distributor reply form. When the form is received, and it is noted that affected product is on hand, Integra Customer Care will contact distributors and provide a Return Material Authorization (RMA) number and directions to return the affected product.
- Quantity
- 40 packs (200 units)
- Distribution
- US Nationwide distribution in the states of Arizona (AZ), Arkansas (AR), California (CA), Connecticut (CT), Florida (FL), Kentucky (KY), Louisiana (LA), Michigan (MI), Minnesota (MN), Missouri (MO), North Carolina (NC), North Dakota (ND), Oklahoma (OK), Oregon (OR), Texas (TX), Washington (WA).
- Lot, serial or model codes
- Product ID: 37.550.501. UDI-DI: 10381780530138. Lot Number: 45322.
- 510(k) numbers
- ["K170599"]
- PMA numbers
- not on file
- Product code
- JXG
- Firm FEI
- 3003418325
- Firm city
- Princeton
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28