Skip to the content

FDA Medical Device Recalls (openFDA)

Laserex Systems Inc.: Ellex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for selective laser trabeculoplasty (ST) operations (laser trabeculoplasty for primary open angle glaucoma). In the VAG mode it is intended to be used to perform procedures requiring the rupture of tissue in the eye for Iridotomy and Iridectomy, Posterior capsulotomy and Posterior membranectomy.

Recall number
Z-2172-2017
Recalling firm
Laserex Systems Inc.
Product
Ellex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for selective laser trabeculoplasty (ST) operations (laser trabeculoplasty for primary open angle glaucoma). In the VAG mode it is intended to be used to perform procedures requiring the rupture of tissue in the eye for Iridotomy and Iridectomy, Posterior capsulotomy and Posterior membranectomy.
Status
Terminated
Initiated
2017-04-04
Posted
2017-06-06
Terminated
2017-07-31
Reason
It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.
Root cause
Device Design
Action
Ellex Medical shall, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard. 1. Installation of a spacer/collar to prevent the slit lamp from being lowered to the point where the cable can be crushed, 2. Improvement of cable management, 3. The corrections will be conducted at no cost to the purchaser, and 4. The corrective action will be completed by June 30, 2017. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. For further questions, please call (800) 824-7444.
Quantity
83
Distribution
US Distribution
Lot, serial or model codes
Serial No: TR 0010, TR 0095
510(k) numbers
not on file
PMA numbers
not on file
Product code
GEX
Firm FEI
1000520094
Firm city
EDEN PRAIRIE
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register