Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Neurovascular: Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 2 CM MODEL Number:M0035431520 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Recall number
Z-2169-2016
Recalling firm
Stryker Neurovascular
Product
Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 2 CM MODEL Number:M0035431520 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Status
Terminated
Initiated
2016-06-06
Posted
2016-07-11
Terminated
2017-03-03
Reason
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Root cause
Process change control
Action
Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instructed to complete the attached customer response form and return the completed form to their nominated Stryker Representative.
Quantity
45,302 (8,188 in the U.S.) - total, all model numbers
Distribution
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Lot, serial or model codes
MODEL Number: M0035431520:  Lot Number/Expiration Date: 18784673 31-Jan-19 18723226 30-Nov-18 18686302 30-Nov-18 18637332 31-Oct-18 18627687 31-Oct-18 18569824 31-Aug-18 18453278 30-Jun-18 18391505 31-May-18 18769932 31-Dec-18 18735994 30-Nov-18 18710356 30-Nov-18 18652851 31-Oct-18 18801847 31-Jan-19 18755945 31-Dec-18 18602481 30-Sep-18 18832966 28-Feb-19 18843005 28-Feb-19 18630099 31-Oct-18 18510994 31-Jul-18 18495363 31-Jul-18 18484141 31-Jul-18 18803162 31-Jan-19 18786297 31-Jan-19 18754879 31-Dec-18 18734778 30-Nov-18 18602423 30-Sep-18 18585290 31-Aug-18 18450430 30-Jun-18 18769816 31-Dec-18 18785485 31-Jan-19 18637274 31-Oct-18 18511110 31-Jul-18 18465901 30-Jun-18 18769700 31-Dec-18 18746995 31-Dec-18 18725103 30-Nov-18 18694228 30-Nov-18 18686012 30-Nov-18 18665561 31-Oct-18 18607752 30-Sep-18 18449575 30-Jun-18 18832355 28-Feb-19 18831744 31-Jan-19 18842138 28-Feb-19 18732878 30-Nov-18 18637216 31-Oct-18 18584252 31-Aug-18 18569882 31-Aug-18 18569766 31-Aug-18 18523410 31-Jul-18 19033487 30-Apr-19 18984845 30-Apr-19 18972140 30-Apr-19 18861952 28-Feb-19 18878607 28-Feb-19 19063170 31-May-19 18923191 31-Mar-19 18897527 28-Feb-19 18897185 28-Feb-19 18910290 31-Mar-19 19032199 30-Apr-19 19002213 30-Apr-19 19002155 30-Apr-19 18950942 31-Mar-19 18897698 28-Feb-19 18858983 28-Feb-19 19031329 30-Apr-19 19031909 30-Apr-19 19002271 30-Apr-19 18951058 31-Mar-19 18950884 31-Mar-19 18896843 28-Feb-19 18897014 28-Feb-19 18879010 28-Feb-19 18909942 31-Mar-19 18910638 31-Mar-19 18843872 28-Feb-19 19031619 30-Apr-19 19002329 30-Apr-19 18878204 28-Feb-19 18886556 28-Feb-19 18938172 31-Mar-19 18939594 31-Mar-19 18951000 31-Mar-19 18897356 28-Feb-19 18885142 28-Feb-19 18971589 30-Apr-19 18950826 31-Mar-19 18860449 28-Feb-19.
510(k) numbers
["K113412","K153658"]
PMA numbers
not on file
Product code
HCG
Firm FEI
3008853977
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register