Hologic, Inc: SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.

Recall number
Z-2167-2026
Recalling firm
Hologic, Inc
Product
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.
Status
Open, Classified
Initiated
2026-04-08
Posted
2026-05-12
Terminated
not on file
Reason
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Root cause
Device Design
Action
On April 8, 2026 URGENT FIELD SAFETY NOTICE letters were sent to customers. Customers were informed of implementing an updated recurring maintenance inspection for detecting loosened, missing, or broken bolts within the VTA. Hologic Service will reach out to schedule your service appointment and perform the required inspection. Following the initial inspection, recurring inspections are required to be conducted at least every 6 months to ensure continued safe/proper operation of the system. Future inspections will consist of torque and visual confirmation that all bolts are present and that there is no indication of loosening. Actions: You may continue to use your 3Dimensions or Selenia Dimensions for its intended use. If at any time during system operation, you observe unexpected C-Arm movement, please discontinue use and contact Hologic s Technical Support immediately. Report suspected device complaints and/or adverse events to Hologic and the FDA (contact details below). After initial inspection by Hologic, it is required to have the inspection performed at least every 6 months as well as the other defined maintenance and service activities which are critical for continued safe and effective operation of the Selenia Dimensions or 3Dimensions system. Forward this notice to anyone in your facility that needs to be informed. Post a copy of this notice in a visible area near the affected devices where operators can view for awareness and please keep a copy for your records.
Quantity
4,305 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
"Model No.; UDI": [SDA-SYS-3000-2D-HTC;15420045510616], [SDA-SYS-3000-3D;15420045517783], [SDM-SYS-6000-2D;15420045510753], [SDM-SYS-6000-3D;15420045510760], [SDM-00001-2D;15420045510623], [SDM-00001-3D;15420045510630], [SDM-00001-M2D;15420045510647], [SDM-00001-M3D;15420045510654], [SDM-05000-2A2;15420045510661], [SDM-05000-2A3;15420045510678], [SDM-05000-2AC;15420045510685], [SDM-05000-2D2;15420045510692], [SDM-05000-2D3;15420045510708], [SDM-05000-2DC;15420045510715], [SDM-05000-3D2;15420045510722], [SDM-05000-3D3;15420045510739], [SDM-05000-3DC;15420045510746], [SDM-SYS-9000-2D;15420045510777], [SDM-SYS-9000-3D;15420045510784]; All Serial No.
510(k) numbers
not on file
PMA numbers
["P010025"]
Product code
MUE
Firm FEI
1000120743
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register