Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories: Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.

Recall number
Z-2167-2017
Recalling firm
Randox Laboratories
Product
Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.
Status
Terminated
Initiated
2017-04-27
Posted
2017-05-26
Terminated
2017-08-14
Reason
According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
Root cause
Error in labeling
Action
Randox Laboratories will contact Customers directly by e-mail and will be followed up with once every 2 weeks over a period of 6 weeks. There are no plans in place to follow up with customers after the 6 week period. For further questions, please call (304) 728-2890.
Quantity
198 kits
Distribution
Worldwide Distribution - US Distribution to Puerto Rico and to the countries of : Canada Kuwait Chile Poland China Romania France Saudi Arabia Guatemala Slovenia Iran UAE India UK Italy Uruguay Iraq
Lot, serial or model codes
Catalogue number UC5075 Batch number 907UC
510(k) numbers
["K110904"]
PMA numbers
not on file
Product code
JJY
Firm FEI
3003059701
Firm city
Kearneysville
Firm state
WV
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register